Specifications

SECTION IV
MAINTENANCE / SERVICE
NOTE
The assembled level sensor rod must be installed
from the inside of the pressure vessel.
(3) Install assembled level sensor rod (4) in fit-
ting (6).
EQUIPMENT ALERT
The end of the silicone tube must extend
past nut (5) after nut is installed. If it is
not, the level sensor rod will not function properly
because the terminal will be in contact with nut (5).
(4) Install and crimp crimp (10) on silicone
tube (9).
(5) Install nut (5).
(6) Install terminal (2) on level sensor rod (4) and
secure using nut (3).
(7) Connect wire (1) to terminal (2).
(8) Install wire tray rack and tray plate (Refer to
para 4.17).
(9) Install back panel (Refer to para 4.5).
4.11 Vent Solenoid Removal /
Installation
NOTE
Steps will vary slightly for M9D / M11D units. Use
this procedure as a guide.
A. Removal
NOTE
Units prior to serial number CZ1110/OM1000 have a
manifold assembly which secures the vent solenoid.
Refer to the parts list to remove the vent solenoid on
these sterilizers.
(1) Drain water from condensing tank.
(2) Remove back panel (Refer to para 4.5).
(3) Tag and disconnect two wires (1, Figure 4-10)
from terminals (3).
(4) Tag and disconnect two wires (2) from terminals
(4).
(5) Disconnect three tubes (5 thru 7) from elbows (8
thru 10).
(6) Remove two screws (11) from bottom of
manifold assembly (12).
NOTE
Spacer (12A) is only on newer units.
(7) Remove manifold assembly (12) and spacer
(12A) from base (13).
(8) Hold screw (1, Figure 4-11) and loosen nut (2).
Turn terminals out of the way.
(9) Remove assembled vent solenoid (3) and elbow
(4) from tee (5).
(10) Remove elbows (4 and 6) from vent sole-
noid (3).
MA2055-01
1
2
3
4
5
7
8
9
6
10
Figure 4-9. Pressure Vessel Water Level
Sensor Removal / Installation