Specifications

SECTION IV
MAINTENANCE / SERVICE
4.5 Back Panel Removal / Installation
A. Removal
(1) Disconnect power cord from sterilizer.
(2) Remove right hand side panel (Refer to
para 4.3).
(3) Remove left hand side panel (Refer to
para 4.4).
(4) Remove four screws (1, Figure 4-4) from back
panel (2).
(5) Partially remove back panel (2) by simulta-
neously pulling downward and outward on back
panel.
EQUIPMENT ALERT
Ensure that wires are completely discon-
nected before attempting to remove back
panel. Failure to do so could result in damage to
sterilizer.
NOTE
Units prior to serial number CZ1110/OM1000 have a
fuse holder attached to the back panel. On these
sterilizers, tag and disconnect two wires from the
receptacle and one wire from the fuse holder instead
of performing step 6.
(6) On older units, tag and disconnect three wires
(3) from terminals of receptacle (4) and remove
back panel (2) from sterilizer. On newer units,
separate back panel (2) from unit as far as
possible and lay it on its side.
B. Installation
NOTE
Units prior to serial number CZ1110/OM1000 have a
fuse holder attached to the back panel. On these
sterilizers, remove tags and connect two wires to the
receptacle and one wire to the fuse holder instead of
performing step 1.
(1) On older units, remove tags and connect three
wires (3) to terminals of receptacle (4).
(2) Position flap of back panel (2) behind top cover
(5) and secure back panel in place using four
screws (1).
(3) Install left hand side panel (Refer to para 4.4).
(4) Install right hand side panel (Refer to para 4.3).
4.6 Base Inspection Cover Removal /
Installation
A. Removal
(1) Disconnect power cord from outlet.
(2) Drain water from condensing tank.
(3) Turn sterilizer onto its back.
MA2051-01
1
2
3
4
5
FLAP
Figure 4-4. Back Panel Removal / Installation
© Midmark Corporation 2002 SF-1827 Page 4-3 Printed in U.S.A.
Return To Table Of Contents