Specifications
SECTION IV
MAINTENANCE / SERVICE
Figure 4-3. Left Hand Side Panel
Removal / Installation
(3) Remove left screw (1, Figure 4-3) from steam
block (2).
(4) Remove two screws (3) from top cover (4).
(5) Lift up the left hand corner of top cover (4), pull
outward and down on the top edge of the left
hand side panel (5), and remove left hand side
panel from sterilizer.
B. Installation
(1) Insert two tabs of left hand side panel (5) into
two slots of base (6).
(2) Lift up the left hand corner of top cover (4), raise
the top edge of left hand side panel (5) into
position, and secure using two screws (3).
(3) Install left screw (1) on steam block (2).
(4) Close sterilizer door.
MA2050-01
1
2
3 4
5
6
SLOT
TAB
© Midmark Corporation 2002 SF-1827 Page 4-2 Printed in U.S.A.
(3) Pull outward and down on the top edge of the
right hand side panel (3) and remove right hand
side panel from sterilizer.
B. Installation
(1) Insert two tabs of right hand side panel (3) into
two slots of base (4).
(2) Raise top edge of right hand side panel (3) into
position and secure using two screws (1).
4.4 Left Hand Side Panel Removal / Instal-
lation
A. Removal
(1) Disconnect power cord from the outlet.
(2) Open sterilizer door.
1
2
3
4
SLOT
MA2049-01
TAB
Figure 4-2. Right Hand Side Panel
Removal / Installation
Return To Table Of Contents