4.2 Top Cover Removal / Installation
A. Removal
(1) Disconnect power cord from the outlet.
(2) If the sterilizer contains a printer, remove printer
(Refer to Operation Manual).
(3) Open sterilizer door.
(4) Remove two screws (1, Figure 4-1) and steam
block (2) from top cover (3).
(5) Remove four screws (4) and two screws (5)
from top cover (3).
(6) On units which apply, carefully break front of
top cover (3) loose from silicone sealant.
EQUIPMENT ALERT
Lift top cover carefully and slowly. Rib-
bon connector is still connected to display
PC board and any excess pressure exerted on it
could result in a damaged ribbon connector or display
PC board.
(7) Lift rear of top cover (3) and disconnect ribbon
connector (6) from display PC board (7A) or
control PC board (7B) (depending on whether
ribbon connector (6) is part of display PC board
or is a separate component).
(8) Remove top cover (3) from sterilizer.
B. Installation
(1) Coat mating surfaces of front panel lip and top
cover (3) with silicone sealant.
(2) Position top cover (3) over sterilizer, with rear of
top cover raised, and connect ribbon connector
(6) to display PC board (7A) or control PC
board (7B).
(3) Install top cover (3) on sterilizer and secure
using two screws (5) and four screws (4).
(4) Install steam block (2) on top cover (3) and
secure using two screws (1).
(5) Close sterilizer door.
(6) If the sterilizer contains a printer, install printer
(Refer to Operation Manual).
4.3 Right Hand Side Panel Removal /
Installation
A. Removal
(1) Disconnect power cord from the outlet.
(2) Remove two screws (1, Figure 4-2) from top
cover (2).
SECTION IV
MAINTENANCE / SERVICE INSTRUCTIONS