Go To Table Of Contents M9 / M9D M11 / M11D To purchase a printed copy of this manual, click on the "Place Order" button below. Place Order -001 thru -019 Self-Contained Steam Sterilizer Serial Number Prefixes: M9 [CZ, DA, DB, DX, DY, FK, LA, FL, OM] M9D [FF, FG] M 11 [ES, ET, FP, FR, GB. NP] M11D [GC, GD] NO LO Som PRO NGE R D e U IN ser be C TIO vice ava N ilab p ar ts le f or t ma yn his ot pro duc t! M9/D M11/D -001 thru -019 FOR USE BY MIDMARK TRAINED TECHNICIANS ONLY SF-1827 Part No.
Go To Table Of Contents
TABLE OF CONTENTS Section/Paragraph Page IMPORTANT INSTRUCTIONS General Safety Instructions......................................... iii Safety Alert Symbols .................................................. iii Warranty Instructions .................................................. iii SECTION I GENERAL INFORMATION 1.1 Model / Serial Number Location................... 1-1 1.2 Specifications .............................................. 1-1 1.3 Description of Normal Operation .................. 1-2 1.
TABLE OF CONTENTS Section/Paragraph Page SECTION VI PARTS LIST - continued NOTE: The Parts List in this manual is divided into two sections [M9 / M9D & M11/M11D]. Verify model and refer to appropriate section. M11 / M11D Parts List M11 Component / Page Cross-Reference ................................... 6-17 M11D Component / Page Cross-Reference .................................... 6-18 Main Enclosure Components.................... 6-19* Plumbing and Sensor Components ..........
Return To Table Of Contents IMPORTANT INSTRUCTIONS General Safety Instructions NOTE Safety First: The primary concern of Midmark Corporation is that this sterilizer is maintained with the safety of the patient and staff in mind. To assure that services and repairs are completed safely and correctly, proceed as follows: (1) Read this entire manual before performing any services or repairs on this sterilizer.
Return To Table Of Contents
Return To Table Of Contents SECTION I GENERAL INFORMATION SECTION I GENERAL INFORMATION 1.1 Model / Serial Number Location MODEL NUMBER A D SERIAL NUMBER C B MA2071-01 Figure 1-1. Model / Serial Number Location Figure 1-2. Dimensions 1.2 Specifications Table 1-1. Specifications Description Data Description Data M9 / M9D Dimensions: Front Height (A, Figure 1-2) .............. 16.4 in (41.8 cm) (printer adds approx. 0.5 in) Width (B) .............................................. 15 in (38.
Return To Table Of Contents SECTION I GENERAL INFORMATION 1.3 Description of Normal Operation NOTE (Figure 1-3 shows Display Panel Buttons & Lamps) The following information applies to all models of the M9 / M9D / M11 / M11D included in this manual. (Figure 1-4 shows location of Major Components) Plug Power Cord Into Outlet Display: E001 (Error lamp flashes) Table 1-1. Specifications - continued Description Data M9 / M9D / M11 / M11D Electrical Requirements: 100 VAC Unit ............................
Return To Table Of Contents SECTION I GENERAL INFORMATION TEMP / TIME DISPLAY LAMP LAMP TEMP (°F) \ TIME (MIN : SECONDS) FILLING STERILIZING DRYING COMPLETE WATER LOW DOOR AJAR ERROR UNWRAPPED POUCHES LIQUIDS PACKS START STOP ON/STANDBY PRESSURE (PSI) PRESSURE DISPLAY UNWRAPPED BUTTON POUCHES BUTTON ON / STANDBY BUTTON PACKS BUTTON START BUTTON STOP BUTTON LIQUIDS BUTTON MA207702 Figure 1-3. Display Panel Button / Lamp Locations Sterilization Cycle - continued 3.
Return To Table Of Contents SECTION I GENERAL INFORMATION DISPLAY PC BOARD HEATING ELEMENT PRESSURE VESSEL PRESSURE RELIEF VALUE BELLOWS ASSEMBLY DOOR SWITCH CONDENSING TANK ASSEMBLY PULSE SOLENOID (M9 / M11only) CONDENSING TANK WATER LEVEL SENSOR (early M9 only) TEMPERATURE SENSOR ASSEMBLY CONTROL PC BOARD PRESSURE VESSEL WATER LEVEL SENSOR THERMOSTAT VENT SOLENOID FILL SOLENOID MA207604 Figure 1-4. Major Components 1.4 Ordering Parts 1.
Return To Table Of Contents SECTION II TESTING AND TROUBLESHOOTING SECTION II TESTING AND TROUBLESHOOTING 2.1 Operational Test 2.2 Troubleshooting Procedures To effectively diagnose a malfunction of the sterilizer, perform the Pressure / Temperature Potentiometers Adjustment procedure. Refer to para 4.24. If a malfunction is detected, locate the Problem & Symptom in Table 2-1. Perform the Check for each Probable Cause indicated. Follow instructions in the Correction column as they apply. Table 2-1.
Return To Table Of Contents SECTION II TESTING AND TROUBLESHOOTING Table 2-1. Troubleshooting Guide - continued Problem Error code 5 [Display pressure exceeded 35 psi (241 kPa)] Error code 6 [Display temperature exceeded 277°F (136°C)] Error code 7 [Display pressure dropped below 24.
Return To Table Of Contents SECTION II TESTING AND TROUBLESHOOTING Table 2-1. Troubleshooting Guide - continued Problem Error code 8 - M9 & M11 only [Door switch did not change status after door solenoid was activated. . Error code 9 [Pressure exceeded 0.9 psi (6.2 kPa) during Dry cycle] Symptom Probable Cause Check Correction Perform continuity check on door switch. Check for loose connections. Replace door switch. Refer to para 4.14. Door opens at end of cycle. Door switch malfunctioning.
Return To Table Of Contents SECTION II TESTING AND TROUBLESHOOTING Table 2-1. Troubleshooting Guide - continued Problem Symptom Check Correction Control PC board has had a software / hardware failure. DIsconnect power; then check for disconnected wiring in sterilizer or facility. Connect power & run additional cycle. If error code persists, replace control PC board. Refer to para 4.13. Sterilizer does not complete (Software interrupt error) cycle.
Return To Table Of Contents SECTION II TESTING AND TROUBLESHOOTING Table 2-1. Troubleshooting Guide - continued Problem Sterilizer has no power. Fill cycle malfunction. Symptom No response when ON/STANDBY is pushed (no lamps illuminate). Unit skips fill cycle. Chamber fills very slowly. Chamber overfills with water. © Midmark Corporation 2002 SF-1827 Probable Cause Check Correction Sterilizer is not plugged in. Check that power cord is plugged in to sterilizer and outlet.
Return To Table Of Contents SECTION II TESTING AND TROUBLESHOOTING Table 2-1. Troubleshooting Guide - continued Problem Door does not open after vent cycle. Symptom Pulse solenoid energizes, but door does not open. Probable Cause Check Correction Latch lever binding. Operate latch lever by hand to check for resistance. Clean / lubricate latch lever. Door latch binding. Operate door latch by hand to check for resistance. Clean / lubricate door latch.
Return To Table Of Contents SECTION II TESTING AND TROUBLESHOOTING Table 2-1. Troubleshooting Guide - continued Problem OPEN THE DOOR lamp does not illuminate. Symptom Probable Cause Check Correction OPEN THE DOOR lamp does not illuminate at the end of sterlization cycle. Pressure potentiometer out of adjustment Display PC board check will show zero pressure setting. Refer to para 4.25. Adjust pressure potentiometer. refer to para 4.24. Cycle will not run.
Return To Table Of Contents SECTION II TESTING AND TROUBLESHOOTING 2.3 Areas To Check For Pressure Leaks (Figure 2-1) 1 1a 5 2a 4a 4 3a 6a 6 5a 3 2 MA542300 Table 2-2. Areas To Check For Pressure Leaks Component Check Correction Door Gasket (1, Figure 2-1) Check for water leaking around door. See (1a). Replace door gasket. Refer to para 4.20. Vent Solenoid (2) Check for water leaking thru condensing coil. See (2a). Replace vent solenoid. Refer to para 4.11.
Return To Table Of Contents SECTION III SCHEDULED MAINTENANCE SECTION III SCHEDULED MAINTENANCE 3.1 Scheduled Maintenance Table 3-1 lists inspections and services that should be performed periodically on the sterilizer. Table 3-1.
Return To Table Of Contents SECTION III SCHEDULED MAINTENANCE © Midmark Corporation 2002 SF-1827 Page 3-2 Printed in U.S.A.
Return To Table Of Contents SECTION IV MAINTENANCE / SERVICE SECTION IV MAINTENANCE / SERVICE INSTRUCTIONS (6) On units which apply, carefully break front of top cover (3) loose from silicone sealant. 4.1 Safety Precautions WARNING EQUIPMENT ALERT Always disconnect the power cord from the outlet before removing any of the sterilizer covers/panels or making any repairs to prevent the possibility of electrical shock. Also, drain the sterilizer to prevent spills during repairs.
Return To Table Of Contents SECTION IV MAINTENANCE / SERVICE 1 (3) Remove left screw (1, Figure 4-3) from steam block (2). 2 (4) Remove two screws (3) from top cover (4). 3 (5) Lift up the left hand corner of top cover (4), pull outward and down on the top edge of the left hand side panel (5), and remove left hand side panel from sterilizer. B. Installation (1) Insert two tabs of left hand side panel (5) into two slots of base (6).
Return To Table Of Contents SECTION IV MAINTENANCE / SERVICE 4.5 Back Panel Removal / Installation EQUIPMENT ALERT A. Removal Ensure that wires are completely disconnected before attempting to remove back panel. Failure to do so could result in damage to sterilizer. (1) Disconnect power cord from sterilizer. (2) Remove right hand side panel (Refer to para 4.3). NOTE Units prior to serial number CZ1110/OM1000 have a fuse holder attached to the back panel.
Return To Table Of Contents SECTION IV MAINTENANCE / SERVICE (4) Remove two screws (1, Figure 4-5) and base inspection cover (2) from base (3). B. Installation TUBE HOLE 4 (1) Install base inspection cover (2) on base (3) and secure using two screws (1). 5 (2) Turn sterilizer upright. (3) Refill condensing tank with distilled water. 3 1 2 1 2 3 MA2052-01 MA2053-01 Figure 4-5. Base Inspection Cover Removal / Installation 4.7 Bellows Assembly Removal / Installation Figure 4-6.
Return To Table Of Contents SECTION IV MAINTENANCE / SERVICE 4.8 Condensing Tank Water Level Sensor Removal / Installation (Applies only to Units with old style Control PC Board with EPROM Version M or Before) NOTE 3 Units with old style control PC board with EPROM version N or after or new style control PC board do not have a condensing tank water level sensor. 6 2 4 A. Removal 7 8 (1) Drain water from condensing tank. 5 (2) Remove back panel (Refer to para 4.5).
Return To Table Of Contents SECTION IV MAINTENANCE / SERVICE 4.9 Temperature Sensor Assembly Removal / Installation (3) Install flex guard tubing (3) on three wire harnesses (1). A. Removal (4) Connect three wire harnesses (1) to three terminals (2). (1) Remove back panel (Refer to para 4.5). (5) Install back panel (Refer to para 4.5). NOTE 4.
Return To Table Of Contents SECTION IV MAINTENANCE / SERVICE 4.11 7 4 NOTE 6 8 Vent Solenoid Removal / Installation Steps will vary slightly for M9D / M11D units. Use this procedure as a guide. A. Removal 9 NOTE Units prior to serial number CZ1110/OM1000 have a manifold assembly which secures the vent solenoid. Refer to the parts list to remove the vent solenoid on these sterilizers. 2 10 (1) Drain water from condensing tank. 5 3 (2) Remove back panel (Refer to para 4.5).
Return To Table Of Contents SECTION IV MAINTENANCE / SERVICE EQUIPMENT ALERT 12 3 When installing the vent solenoid, the side of the solenoid marked IN needs to be connected to the pressure line. Reversing vent solenoid will cause system failure. 7 1 10 (7) Install spacer (12A, Figure 4-10) and manifold assembly (12) on base (13) and secure using two screws (11). 6 5 9 (8) Connect three tubes (5 thru 7) to elbows (8 thru 10). (9) Connect two wires (2) to terminals (4).
Return To Table Of Contents SECTION IV MAINTENANCE / SERVICE 4.12 Fill Solenoid Removal / Installation NOTE Spacer (12A) is only on newer units. NOTE Steps will vary slightly for M9D / M11D units. Use this procedure as a guide. (7) Remove manifold assembly (12) and spacer (12A) from base (13). A. Removal (8) Hold screw (1, Figure 4-13) and loosen nut (2). Turn terminals out of the way. NOTE (9) Remove assembled vent solenoid (3) and elbow (4) from tee (5).
Return To Table Of Contents SECTION IV MAINTENANCE / SERVICE (6) Hold screw (1) and loosen nut (2). Turn terminals out of the way. (4) Tag and disconnect two wires (5) from terminals (6). (7) Install assembled elbow (4) and vent solenoid (3) on tee (5). (5) Tag and disconnect two wires (7) from terminals (8). (8) Turn terminals of vent solenoid (3) so wires will connect easily to vent solenoid. (6) Disconnect ribbon connector (9) from connector (10). (9) Hold screw (1) and tighten nut (2).
Return To Table Of Contents SECTION IV MAINTENANCE / SERVICE OLD STYLE PC BOARD 10 18 17 9 20 19 2 1 16 15 3 22 4 21 14 13 12 11 8 5 6 7 10 17 20 9 2 19 1 16 18 NEW STYLE PC BOARD 15 3 21 4 13 22 11 14 12 8 6 5 7 MA207505 Figure 4-14. Control PC Board Connections © Midmark Corporation 2002 SF-1827 Page 4-11 Printed in U.S.A.
Return To Table Of Contents SECTION IV MAINTENANCE / SERVICE (3) Install control PC board (3) and control PC board bracket (4) as an assembly on base (5) and secure using two screws (2) and lock- nuts (1). 9 10 (4) Connect heater element wire (21, Figure 4-14) to connector (22). M9 / M11 only (5) Connect two wires (19) on pulse solenoid terminals (20). 4 (6) Connect wire harness (17) to connector (18). 2 6 (7) Connect printer cable (15) to connector (16). (8) Connect hot wire (13) to terminal (14).
Return To Table Of Contents SECTION IV MAINTENANCE / SERVICE 4.14 Door Switch Removal / Installation B. Installation (1) Position door switch (5) in door switch bracket (4). A. Removal NOTE (2) Using 3/32 in. punch, install two roll pins (3) in door switch bracket (4) (Refer to Table 1-2 for special tool). Units prior to serial number CZ1110/OM1000 have a different switch weldment.
Return To Table Of Contents SECTION IV MAINTENANCE / SERVICE B. Installation (1) Slide thermostat (8) into bracket (6) and rotate terminals (3 and 4) to a horizontal position. PLUNGER 4 5 (2) Position bracket (6) against pressure vessel (7) and secure by tightening two nuts (5). 3 (3) Connect wires (1 and 2) to terminals (3 and 4). (4) Install base inspection cover (Refer to para 4.6). (5) Refill condensing tank with distilled water. 7 1 8 2 6 1 4 MA2060-01 Figure 4-17.
Return To Table Of Contents SECTION IV MAINTENANCE / SERVICE 4.17 (4) Install trays. Wire Tray Rack and Tray Plate Removal / Installation (5) Close sterilizer door. A. Removal 4.18 (1) Open sterilizer door. (2) Remove trays. Heating Element and Gasket Removal / Installation A. Removal (3) Pull wire tray rack (1, Figure 4-19) and tray plate (2) from pressure vessel (3). (1) Drain water from condensing tank. (2) Remove base inspection cover (Refer to 4.6). B.
Return To Table Of Contents SECTION IV MAINTENANCE / SERVICE EQUIPMENT ALERT Hold heating element firmly in position while tightening nuts. Otherwise, damage to heating element or improper positioning of heating element may result. Also, spacer assembly must remain above gaskets. If spacer assembly is installed under gasket, leaking will result. 7 9 2 (6) Install two brass washers (5), lock washers (4), and nuts (3) on heating element (8).
Return To Table Of Contents SECTION IV MAINTENANCE / SERVICE 3 GROOVE 1 SLOT 2 2 MA2063-01 Figure 4-21. Filter Removal / Installation FLAP (2) Using rubber hammer, tap filter (1) lightly into pressure vessel (2). MA2064-01 Figure 4-22. Door Gaskets Removal / Installation (3) Install wire tray rack and tray plate (Refer to para 4.17). 4.20 1 4.21 Door Gaskets Removal / Installation Display PC Board Removal / Installation A. Removal A. Removal (1) Remove top cover (Refer to para 4.2).
Return To Table Of Contents SECTION IV MAINTENANCE / SERVICE 3 1 2 2 1 4 MA2066-01 Figure 4-24. Condensing Tank Assembly Fill Tube Disconnection / Connection NOTE MA2065-01 Figure 4-23. Display PC Board Removal / Installation 4.22 The following step applies only to units with old style control PC board with EPROM version M or before. Units with old style control PC board with EPROM version N or after or new style control PC board do not have wire (4) and terminal (5).
Return To Table Of Contents SECTION IV MAINTENANCE / SERVICE NOTE 4.23 The following step applies only to units with old style control PC board with EPROM version M or before. Units with old style control PC board with EPROM version N or after or new style control PC board do not have wire (4) and terminal (5). Pressure Relief Valve Removal / Installation A. Removal (1) Remove condensing tank assembly (Refer to para 4.22). (3) Connect wire (4) to terminal (5).
Return To Table Of Contents SECTION IV MAINTENANCE / SERVICE 4.24 Pressure / Temperature Potentiometers Adjustments NOTE After the following step is performed, the display PC board lamps will individually illuminate and then extinguish one at a time in a left to right sequence. When all the lamps have illuminated and extinguished, proceed with step 4. EQUIPMENT ALERT The pressure zero, pressure range, and temperature potentiometers cannot be adjusted separately.
Return To Table Of Contents SECTION IV MAINTENANCE / SERVICE (5) Turn adjusting screw of pressure zero potentiometer (1, Figure 4-28) in the counter-clockwise direction until the PRESSURE (PSI) display reads 0.0 psi (0.0 kPa). (6) Turn adjusting screw of pressure zero potentiometer (1) in the clockwise direction until the PRESSURE (PSI) display reads 0.1 psi (1 kPa).
Return To Table Of Contents SECTION IV MAINTENANCE / SERVICE O F F 1 2 3 4 A R3 2 U1 NEW STYLE PC BOARD 3 2 B ADJUSTING SCREWS 2 1 3 1 B A 3 ADJUSTING SCREWS OLD STYLE PC BOARD ADJUSTING SCREW 2 MA407700 Figure 4-28. Potentiometer Adjustments © Midmark Corporation 2002 SF-1827 Page 4-22 Printed in U.S.A.
Return To Table Of Contents SECTION IV MAINTENANCE / SERVICE 4.25 2 (1) Disconnect power cord from the sterilizer. 1 TUBE Display PC Board Lamp / Display / Button Check (2) Press and hold down the LIQUIDS and PACKS buttons simultaneously, while connecting the power cord to the sterilizer (See Figure 4-30). Release the LIQUIDS and PACKS buttons. (3) Observe the display PC board lamps. In a left to right sequence, each lamp on the display PC board should individually illuminate and then extinguish.
Return To Table Of Contents SECTION IV MAINTENANCE / SERVICE TEMP / TIME DISPLAY LAMP LAMP TEMP (°F) \ TIME (MIN : SECONDS) PRESSURE (PSI) FILLING STERILIZING DRYING COMPLETE WATER LOW DOOR AJAR ERROR UNWRAPPED POUCHES LIQUIDS PACKS START STOP ON/STANDBY PRESSURE DISPLAY UNWRAPPED BUTTON POUCHES BUTTON ON / STANDBY BUTTON PACKS BUTTON START BUTTON STOP BUTTON LIQUIDS BUTTON MA207701 Figure 4-30. Display PC Board Lamp / Display / Button Check (M9 / M11 shown) 4.
Return To Table Of Contents SECTION IV MAINTENANCE / SERVICE SETTING #1 REPEAT INTERVAL 1 SETTING #3 REPEAT INTERVAL WIRE RING O F F O F F 1 2 3 4 1 2 3 4 2 #3 SWITCH DEPRESSED TO RIGHT (ON) #4 SWITCH DEPRESSED TO RIGHT (ON) #3 SWITCH DEPRESSED TO LEFT (OFF) #4 SWITCH DEPRESSED TO LEFT (OFF) SETTING #2 REPEAT INTERVAL SETTING #4 REPEAT INTERVAL MA2072-01 Figure 4-31. Pressure Relief Valve Check 4.
Return To Table Of Contents SECTION IV MAINTENANCE / SERVICE 4.28 (4) Press and hold STOP & POUCHES buttons (see Figure 4-33) while plugging in power cord. M9 / M11 Door Opening Test NOTE (5) Press and release all buttons from left to right across display panel. This test applies only to units with newer style PC boards. (1) Fill chamber with water. [To do this, start any cycle. After chamber has filled with water, the Sterilizing LED will illuminate]. (6) Close door.
Return To Table Of Contents SECTION V SCHEMATICS AND DIAGRAMS SECTION V SCHEMATICS AND DIAGRAMS 5.1 Wiring Diagrams Figure 5-1 (sheets 1 thru 3) illustrate the wiring connections for the M9 sterilizer. Figure 5-3 (sheets 1 thru 3) illustrate the wiring connections for the M11 sterilizer. Figure 5-2 (sheets 1 & 2) illustrate the wiring connections for the M9D sterilizer. Figure 5-4 (sheets 1 & 2) illustrate the wiring connections for the M11D sterilizer.
Return To Table Of Contents SECTION V SCHEMATICS AND DIAGRAMS Printer Printer Connector (*) NOTE: Some Solenoid Coils may be marked FWR (Full Wave Rectified). Use the Mohms scale to measure these coils. If reading is 0.2 Mohms or above, coil is GOOD.
Return To Table Of Contents SECTION V SCHEMATICS AND DIAGRAMS (*) NOTE: Some Solenoid Coils may be marked FWR (Full Wave Rectified). Use the Mohms scale to measure these coils. If reading is 0.2 Mohms or above, coil is GOOD. If OL is displayed, coil is BAD Printer Printer Connector 4 5 3 1 1 2 3 4 5 2 Display Panel PC Board Temperature Probe Assembly PCB/Printer Harness Blue (Busy) White (Data) Level Sensor Assembly White (Ground) Flex Guard Tubing Green (Ground) Red (Power) Black (+5 V.
Return To Table Of Contents SECTION V SCHEMATICS AND DIAGRAMS (*) NOTE: Some Solenoid Coils may be marked FWR (Full Wave Rectified). Use the Mohms scale to measure these coils. If reading is 0.2 Mohms or above, coil is GOOD. If OL is displayed, coil is BAD Printer Printer Connector 4 5 3 1 1 2 3 4 5 2 Display Panel PC Board Temperature Probe Assembly PCB/Printer Harness Ribbon Cable Connector Blue (Busy) White (Data) N.O. White (Ground) Red (Power) Black (+5 V.
Return To Table Of Contents SECTION V SCHEMATICS AND DIAGRAMS (*) NOTE: Some Solenoid Coils may be marked FWR (Full Wave Rectified). Use the Mohms scale to measure these coils. If reading is 0.2 Mohms or above, coil is GOOD. If OL is displayed, coil is BAD Printer Printer Connector 4 3 5 1 1 2 3 4 5 2 Display Panel PC Board Temperature Probe Assembly PCB/Printer Harness Red (Power) Blue (Busy) White (Data) Level Sensor Assembly White (Ground) Flex Guard Tubing Green (Ground) Black (+5 V.
Return To Table Of Contents SECTION V SCHEMATICS AND DIAGRAMS (*) NOTE: Some Solenoid Coils may be marked FWR (Full Wave Rectified). Use the Mohms scale to measure these coils. If reading is 0.2 Mohms or above, coil is GOOD.
Return To Table Of Contents SECTION V SCHEMATICS AND DIAGRAMS (*) NOTE: Some Solenoid Coils may be marked FWR (Full Wave Rectified). Use the Mohms scale to measure these coils. If reading is 0.2 Mohms or above, coil is GOOD. If OL is displayed, coil is BAD Printer Printer Connector 4 3 5 1 1 2 3 4 5 2 Display Panel PC Board Temperature Probe Assembly PCB/Printer Harness Ribbon Cable Connector Blue (Busy) White (Data) White (Ground) Red (Power) Black (+5 V.
Return To Table Of Contents SECTION V SCHEMATICS AND DIAGRAMS (*) NOTE: Some Solenoid Coils may be marked FWR (Full Wave Rectified). Use the Mohms scale to measure these coils. If reading is 0.2 Mohms or above, coil is GOOD. If OL is displayed, coil is BAD Printer Printer Connector 4 5 3 1 1 2 3 4 5 2 Display Panel PC Board Temperature Probe Assembly PCB/Printer Harness Red (Power) Blue (Busy) White (Data) Level Sensor Assembly White (Ground) Flex Guard Tubing Green (Ground) Black (+5 V.
Return To Table Of Contents SECTION V SCHEMATICS AND DIAGRAMS (*) NOTE: Some Solenoid Coils may be marked FWR (Full Wave Rectified). Use the Mohms scale to measure these coils. If reading is 0.2 Mohms or above, coil is GOOD. If OL is displayed, coil is BAD Printer Printer Connector 4 5 3 1 1 2 3 4 5 2 Display Panel PC Board Temperature Probe Assembly PCB/Printer Harness Ribbon Cable Connector Blue (Busy) White (Data) White (Ground) Red (Power) Black (+5 V.
Return To Table Of Contents SECTION V SCHEMATICS AND DIAGRAMS (*) NOTE: Some Solenoid Coils may be marked FWR (Full Wave Rectified). Use the Mohms scale to measure these coils. If reading is 0.2 Mohms or above, coil is GOOD. If OL is displayed, coil is BAD Printer Printer Connector 4 5 3 1 1 2 3 4 5 2 Display Panel PC Board Temperature Probe Assembly PCB/Printer Harness Blue (Busy) White (Data) N.O. Level Sensor Assembly White (Ground) Flex Guard Tubing Green (Ground) Red (Power) Black (+5 V.
Return To Table Of Contents SECTION V SCHEMATICS AND DIAGRAMS 5.2 Flow Diagram Figure 5-5 illustrates the water, heated water, steam, and vented steam/water flow throughout the sterilizer during a cycle.
Return To Table Of Contents SECTION V SCHEMATICS AND DIAGRAMS 5.3 Pressure / Temperature Chart Table 5-1 is a chart which lists what pressure should be present at particular temperature. Table 5-1. Pressure / Temperature Chart PSI °C °F PSI °C °F 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 100.0 101.9 103.6 105.3 106.9 108.4 109.8 11.3 112.6 113.9 115.2 116.4 117.6 118.8 119.9 121.0 212.0 215.4 218.5 221.5 224.4 227.1 229.6 232.3 234.7 237.0 239.4 241.5 243.7 245.8 247.8 249.
Return To Table Of Contents SECTION VI PARTS LIST M9 / M9D SECTION VI PARTS LIST 6.1 Description of Bullets 6.2 Torque Specifications and Important Assembly Notes If a component has a bullet [ • ] in the Part No. column and the Description column, it is a subcomponent of an assembly in the parts list. Required torque specifications, measurements, and procedures will be noted on the parts illustrations. Adherence to these requirements is essential.
Return To Table Of Contents SECTION VI PARTS LIST M9 / M9D M9 Component / Page Cross-Reference Labels Trays / Rack Display PC Board Refer to: Labels & Decals Refer to: Racks, Trays, Cleaner Refer to: Electrical Components Top Cover Refer to: Main Enclosure Components Pressure Vessel Refer to: Pressure Vessel Components Heating Element Refer to: Electrical Components Pressure Relief Valve Pulse (Door) Solenoid Refer to: Plumbing & Sensor Components Refer to: Electrical Components Bellows Door S
Return To Table Of Contents SECTION VI PARTS LIST M9 / M9D M9D Component / Page Cross-Reference Labels Trays / Rack Display PC Board Refer to: Labels & Decals Refer to: Racks, Trays, Cleaner Refer to: Electrical Components Top Cover Refer to: Main Enclosure Components Pressure Vessel Refer to: Pressure Vessel Components Heating Element Refer to: Electrical Components Pressure Relief Valve Refer to: Plumbing & Sensor Components Door Switch Refer to: Electrical Components Bellows Refer to: Plumbin
Return To Table Of Contents SECTION VI PARTS LIST M9 / M9D Main Enclosure Components 3 1 8 2 4 5 14 4 6 20 14 8 9 10 7 11 11 8 19 8 8 12 14 13 15 16 21 15 22 18 8 23 14 17 7 14 24 22 MA225400 Used on units with Serial Numbers: CZ1000 thru CZ1456 OM1000 thru OM3287 Item 1 2 3 4 5 6 7 8 9 10 11 12 Part No. Description Qty.
Return To Table Of Contents SECTION VI PARTS LIST M9 / M9D Main Enclosure Components 3 1 8 2 4 5 14 4 6 20 14 8 9 10 7 11 26 11 8 8 19 8 12 14 13 15 25 16 15 18 15 22 18 21 8 21 23 9 7 14 17 24 22 MA294400 Used on units with Serial Numbers: CZ1457 thru CZ2753 DY1000 thru DY1149 Item 1 2 3 4 5 6 7 8 9 10 11 12 13 DA1000 thru DA1004 FD1000 thru FD1099 Part No. Description Qty.
Return To Table Of Contents SECTION VI PARTS LIST M9 / M9D Main Enclosure Components 1 8 3 2 4 14 5 4 20 29 7 8 16 19 6 11 10 31 30 28 8 15 12 14 8 15 13 25 27 26 11 18 21 22 18 8 21 23 9 11 7 17 14 24 22 MA2944012 Used on units with Serial Numbers: CZ2754 thru CZ4530 FD1100 thru FD1499 FL1000 thru FL1474 Item 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 DB1150 thru DB1237 DX2132 thru DX2588 FF1000 thru FF1039 FG1000 thru FG1127 OM10597 thru OM14666 Part No. Description Qty.
Return To Table Of Contents SECTION VI PARTS LIST M9 / M9D Main Enclosure Components 1 8 3 2 4 14 5 4 20 28 7 8 16 19 6 11 10 30 29 27 8 15 12 14 8 15 13 26 25 11 18 21 22 18 8 21 23 9 11 7 17 14 24 22 MA294404 Used on units with Serial Numbers: CZ4531, DA1005, DB1238, DX2589, DY1179, FD1500, FF1040, FG1128, FK1750, FL1475, FX1000, FY1000, LA1000 and OM14667 - thru Present Item 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 Part No. Description Qty.
Return To Table Of Contents SECTION VI PARTS LIST M9 / M9D Plumbing & Sensor Components 21 25 20 15 28 34 19 27 18 24 10 26 9 29 22 4 8 17 7 3 26 23 19 16 6 19 2 10 5 1 9 A 8 C 35 30 31 32 11 33 B 36 37 12 14 13 MA201910 Used on units with Serial Numbers: CZ1000 thru CZ1119 OM1000 thru OM1352 Item 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 Part No. Description Qty.
Return To Table Of Contents SECTION VI PARTS LIST M9 / M9D Plumbing & Sensor Components 21 25 20 15 28 34 19 27 18 24 10 26 9 29 22 4 8 17 7 3 26 23 19 16 6 19 2 10 5 1 9 A 8 C 30 35 31 32 11 33 B 36 37 12 14 13 MA201910 Used on units with Serial Numbers: CZ1120 thru CZ1684 DA1000 thru DA1001 DB1000 (one unit) DX1000 thru DX1100 OM1353 thru OM4864 Item 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 Part No.
Return To Table Of Contents SECTION VI PARTS LIST M9 / M9D Plumbing & Sensor Components 21 25 20 15 28 34 19 27 18 24 10 26 9 29 19 16 22 4 8 17 7 3 26 23 6 19 2 10 5 1 9 A 8 C 35 30 31 32 11 33 B 36 37 12 14 13 DX1101 thru DX1170 Item 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 MA201910 Used on units with Serial Numbers: DY1000 thru DY1007 OM4865 thru OM5130 Part No.
Return To Table Of Contents Plumbing & Sensor Components SECTION VI PARTS LIST M9 / M9D 25 21 34 20 15 36 10 24 19 9 26 27 18 8 19 28 35 16 19 29 26 23 19 4 17 3 22 2 1 A C 6 30 31 32 11 33 B 5 7 12 13 14 MA201916 Used on units with Serial Numbers: CZ1685 thru CZ4334 DY1008 thru DY1176 FK1000 thru FK1699 Item 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 Part No.
Return To Table Of Contents SECTION VI PARTS LIST M9 / M9D Plumbing & Sensor Components 28 25 20 19 18 28 21 27 24 10 26 15 29 23 19 16 26 22 9 17 8 4 3 2 1 A 6 30 31 32 11 33 B 12 7 5 14 13 MA201912 Used on units with Serial Numbers: CZ4335 thru CZ5944 FD1400 thru FD1954 FL1475 thru FL1843 Item DB1238 thru DB1262 DX2556 thru DX2883 FF1035 thru FF1177 FG1113 thru FG1290 OM14525 thru OM17468 Part No.
Return To Table Of Contents SECTION VI PARTS LIST M9 / M9D Plumbing & Sensor Components 25 20 21 19 27 23 18 26 24 10 28 26 15 29 19 16 9 8 22 17 4 3 2 1 A 6 C 30 31 32 11 33 B 7 5 12 14 13 MA510809 Used on units with Serial Numbers: CZ5945 thru CZ6530 FF1178 thru FF1264 LA1000 thru LA1023 Item 1 2 3 5 6 7 8 Part No.
Return To Table Of Contents SECTION VI PARTS LIST M9 / M9D Plumbing & Sensor Components 25 20 21 19 27 23 18 26 24 10 28 26 15 29 19 16 9 8 22 17 4 3 2 1 A 6 C 30 31 32 11 33 B 7 5 12 13 14 MA510809 Used on units with Serial Numbers: CZ6531 thru CZ7236 FK2305 thru FK2324 FG1328 thru FG1386 Item DX2879 thru DX2906 DY1179 thru DY1182 FL1949 thru FL1963 LA1024 thru LA1028 OM18881 thru OM20225 Part No.
Return To Table Of Contents SECTION VI PARTS LIST M9 / M9D Plumbing & Sensor Components 24 19 20 18 26 22 17 25 23 9 27 25 14 28 18 15 8 7 21 16 3 2 1 A 5 C 29 30 31 10 32 B 12 6 4 11 13 MA510804 Used on units with Serial Numbers: CZ7237 thru CZ10064 FD2105 thru FD2264 Item 1 2 4 5 6 7 Part No.
Return To Table Of Contents SECTION VI PARTS LIST M9 / M9D Plumbing & Sensor Components 33 24 19 20 18 26 22 17 25 23 9 00 2 -06 27 25 54 - 0 0 Flow 14 Direct ion Wren ch Here 28 18 15 8 7 21 16 3 2 1 A 5 C 29 30 31 10 32 B 12 6 4 11 13 MA510807 Used on units with Serial Numbers: CZ10065, DA1005, DB1283, DX3148, DY1191, FD2265, FF1737, FG1618, FK2325, FL1964, LA1029, OM25096 - thru Present Item Part No.
Return To Table Of Contents SECTION VI PARTS LIST M9 / M9D Pressure Vessel Components 21 22 20 24 2 23 3 26 1 4 30 36 5 3 27 28 6 29 7 39 11 34 9 33 25 10 14 13 15 8 38 31 12 35 16 17 37 WARNING 32 Do not attempt to adjust, modify, or alter in any manner, any part of the pressure vessel. Serious injury and/or damage to the unit could result.
Return To Table Of Contents SECTION VI PARTS LIST M9 / M9D Pressure Vessel Components 21 22 20 24 2 23 3 26 1 4 30 36 5 3 27 28 6 7 29 39 11 34 9 33 25 10 14 13 15 8 38 31 12 35 16 17 37 32 WARNING 18 Do not attempt to adjust, modify, or alter in any manner, any part of the pressure vessel. Serious injury and/or damage to the unit could result.
Return To Table Of Contents M9 Manifold Components SECTION VI PARTS LIST M9 / M9D 1 2 4 A 3 5 7 6 C B 8 MA202102 Used on units with Serial Number CZ1000 thru CZ1109 Item 1 2 3 4 Part No. Description Qty. 014-0199-00 014-0200-00 014-0198-00 040-0010-53 Solenoid Valve [Vent] ............................ Solenoid Valve [Fill] ............................... Manifold ................................................. Screw ...................................................
Return To Table Of Contents SECTION VI PARTS LIST M9 / M9D M9 Manifold Components A 4 3 5 2 7 6 C 8 9 B 1 NOTE When replacing components in the solenoid, examine the existing valve to determine if it was Full Wave Rectified (if so, will be marked FWR). MA202205 Used on units with Serial Numbers: CZ1110 thru CZ7236 DY1000 thru DY1182 Item Part No. Description DA1000 thru DA1006 FD1000 thru FD2104 LA1000 thru LA1028 Qty.
Return To Table Of Contents SECTION VI PARTS LIST M9 / M9D M9 Manifold Components A 4 3 5 2 7 6 C 8 9 B 1 NOTE When replacing components in the solenoid, examine the existing valve to determine if it was Full Wave Rectified (if so, will be marked FWR). MA510701 Used on units with Serial Numbers: CZ7237, DA1007, DB1273, DX2907, DY1183, FD2105, FK2325, FL1964, LA1029, and OM20226 - thru Present Item Part No. Description Qty.
Return To Table Of Contents SECTION VI PARTS LIST M9 / M9D M9D Manifold Components A 1 4 5 6 2 3 8 7 C 10 9 B 11 NOTE When replacing components in the solenoid, examine the existing valve to determine if it was Full Wave Rectified (if so, will be marked FWR). MA442700 Used on Units with Serial Numbers: FF1000 thru FF1290 FG1000 thru FG1386 Item 1 2 3 4 5 6 7 Part No. Description 014-0183-01 Male Connector ..................................... 1 Reducer Fitting ............................
Return To Table Of Contents SECTION VI PARTS LIST M9 / M9D M9D Manifold Components A 1 4 5 6 2 3 8 7 C 10 B 9 11 NOTE When replacing components in the solenoid, examine the existing valve to determine if it was Full Wave Rectified (if so, will be marked FWR). MA511100 Used on Units with Serial Numbers: FF1291, FG1387 - thru Present Item 1 2 3 4 5 6 7 Part No. 014-0183-01 002-0519-00 • 014-0275-02 014-0194-00 014-0276-00 014-0234-00 014-0184-01 Description Qty. Male Connector ..................
Return To Table Of Contents SECTION VI PARTS LIST M9 / M9D Electrical Components 4 3 9 1 3 10 6 2 1 1 13 8 19 7 12 23 5 15 22 14 24 20 18 26 17 21 16 34 25 30 27 33 28 31 32 29 MA202301 CZ1000 thru CZ1684 DX1000 thru DX1170 Item 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 Part No.
Return To Table Of Contents Electrical Components SECTION VI PARTS LIST M9 / M9D 4 3 9 1 3 6 2 10 11 13 8 19 7 23 14 12 15 22 5 14 23 24 18 26 17 21 16 20 25 30 27 33 28 31 32 29 MA2023-02 CZ1685 thru CZ2456 DX1171 thru DX1969 Item 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 Part No.
Return To Table Of Contents SECTION VI PARTS LIST M9 / M9D Electrical Components 2 10 9 8 1 11 13 7 19 12 23 15 22 5 14 24 18 26 17 21 16 20 25 30 27 6 28 3 4 29 MA202309 Used on units with Serial Numbers: CZ2457 thru CZ4590 Item 1 2 3 4 5 6 7 8 9 10 Part No.
Return To Table Of Contents SECTION VI PARTS LIST M9 / M9D Electrical Components 2 10 9 8 1 11 7 31 18 12 22 13 14 21 5 13 23 17 25 16 20 15 30 19 24 29 1 26 6 27 3 4 28 MA376103 Used on Units with Serial Numbers: CZ4591, DA1005, FK1000, FL1000, Item 1 2 3 4 5 6 7 8 9 10 Part No.
Return To Table Of Contents SECTION VI PARTS LIST M9 / M9D Electrical Components 10 9 2 8 11 7 1 27 18 22 13 14 21 5 22 23 17 25 16 20 15 24 6 26 19 3 4 12 MA442901 Used on Units with Serial Numbers: FF1055 and FG1138 - thru Present Item Part No. 1 2 3 4 5 6 7 8 9 10 11 015-1137-00 002-0370-00 057-0425-00 040-0010-75 H225617 040-0006-04 002-0369-02 053-0454-00 061-0531-00 12 13 14 H97949 040-0010-46 015-0657-00 Description Qty. Cordset (Refer to Wiring Diagram) ........
Return To Table Of Contents SECTION VI PARTS LIST M9 / M9D M9 Control PC Board 1 2 4 3 MA2220-01 CZ1000 thru CZ2456 DX1170 thru DX1969 Item Part No. Description 002-0434-00 Control PC Board {115 V Domestic} (Includes Item 1thru 3) .......................... 1 Control PC Board {230 V Export} (Includes Item 1 thru 3) ......................... 1 Control PC Board {100 V Export} (Includes Item 1 thru 4) ......................... 1 • M9 EPROM ......................................
Return To Table Of Contents SECTION VI PARTS LIST M9 / M9D M9 Control PC Board 1 2 3 MA281100 Used on Units with Serial Numbers: FD1000 thru FD1399 Item 1 Part No. Description 002-0434-04 Control PC Board {230 V. French Export} (Includes Item 1thru 3) .......................... 1 • M9 EPROM ...................................... Ref • Qty. Item 2 3 Part No. Description • 015-0346-08 • 015-0346-11 • Printer Fuse (3 amp / 250 V.) ..............
Return To Table Of Contents SECTION VI PARTS LIST M9 / M9D M9 Control PC Board 1 2 3 MA281100 Used on units with Serial Numbers: CZ2457, Item DA1005, FL1000, DB1150, DX1970, DY1139, FD1400, LA1000, and OM9449 - thru Present Part No. Description 002-0434-03 Control PC Board {115 V Domestic} (Includes Item 1thru 3) .......................... Control PC Board {230 V French Export} (Includes Item 1 thru 3) ......................... Control PC Board {230 V Export} (Includes Item 1 thru 3) ............
Return To Table Of Contents SECTION VI PARTS LIST M9 / M9D M9D Control PC Board 1 2 3 MA443000 Item 1 2 Part No. Description 002-0518-00 Control PC Board {115 V } (Includes Item 1thru 3 ) ......................... 1 • M9 EPROM ...................................... Ref • Printer Fuse (3 amp/250 V) ................. 1 • • 015-0346-08 Qty. Item 3 Part No. Description Qty. • 015-0346-07 • Line Fuse (15 amp/250 V) {Used on 115 V Units} ...................................
Return To Table Of Contents M9 Labels and Decals SECTION VI PARTS LIST M9 / M9D 5 12 11 4 6 9 3 7 10 13 1 2 8 MA202513 Item 1 2 3 4 5 6 Part No. Description 061-0203-00 061-0516-00 061-0600-00 Label - Electrical Shock (English) .......... 1 Label -Electrical Shock (French) ........... 1 Label -Electrical Shock (Portuguese) ..... 1 Label - Serial Number (Front) ................ 1 Label - Patent ........................................ 1 Label - Serial Number (Rear) .................
Return To Table Of Contents SECTION VI PARTS LIST M9 / M9D M9D Labels and Decals 5 4 9 6 3 10 7 12 1 2 8 MA443101 Item 1 2 3 4 5 6 7 8 Part No. Description 061-0203-00 Label - Electrical Shock ......................... Label - Patent (Front) ............................ Label - Patent (Rear) ............................. Label - Serial Number ............................ Label - Warning, P.R. Valve ................... Label - Warning ..................................... Label - Hot Surface ..
Return To Table Of Contents Printer Components SECTION VI PARTS LIST M9 / M9D 3 4 2 5 6 OF F ON /F D EE 1 MA203302 Item 1 2 3 4 Part No. Description Qty. 9A155001 • 053-0506-00 • 060-0008-00 • 053-0507-00 Accessory Printer (Incl. Items 2 thru 6) . • Arbor ................................................... • Paper Roll ........................................... • Cover .................................................. 1 1 1 1 Item 5 6 7 Part No. Description Qty.
Return To Table Of Contents SECTION VI PARTS LIST M9 / M9D Racks, Trays, and Cleaner 7 1 2 6 4 3 Cleaner 5 2 MA203201 Used on Units with Serial Numbers: CZ1000 thru CZ4334 DY1000 thru DY1176 FK1000 thru FK1699 Item 1 2 3 4 5 DA1000 thru DA1004 FD1000 thru FD1399 FL1000 thru FL1474 Part No. Description 030-0710-01 002-0253-00 002-0374-00 050-1773-00 002-0269-00 Wire Tray Rack ..................................... 1 5" Tray .................................................. 2 7" Tray ...........
Return To Table Of Contents Racks, Trays, and Cleaner SECTION VI PARTS LIST M9 / M9D 7 1 2 6 4 3 Cleaner 5 2 MA376300 Used on units with Serial Numbers: CZ4335, DA1005, DB1238, DX2556, DY1177, FD1400, FF1035, FG1113, FK1700, FL1475, LA1000, OM14525 - thru Present Item 1 2 3 4 5 Part No. Description 050-3691-00 002-0253-00 002-0374-00 050-3750-00 002-0269-00 Wire Tray Rack ..................................... 1 5" Tray .................................................. 2 7" Tray .............
Return To Table Of Contents SECTION VI PARTS LIST M9 / M9D Packaging 7 5 6 3 4 3 2 1 MA202406 Item 1 Part No. Description 003-1031-00 M9 Installation & Operation Manual (English) .................................... M9D Installation & Operation Manual (English) ..................................... M9 Installation & Operation Manual (French) .................................... M9 Installation & Operation Manual (Polish) ..................................... Slotted Carton .....................
Return To Table Of Contents SECTION VI PARTS LIST M11 / M11D SECTION VI PARTS LIST Attention! The Parts Lists in this manual are separated as shown below. Verify model number, then refer to appropriate pages. M9 / M9D .......................... Pages 6-2 thru 6-16 M11 / M11D ..................... Pages 6-17 thru 6-31 © Midmark Corporation 2002 SF-1827 Printed in U.S.A.
Return To Table Of Contents SECTION VI PARTS LIST M11 / M11D M11 Component / Page Cross-Reference Labels Trays / Rack Display PC Board Refer to: Labels & Decals Refer to: Racks, Trays, Cleaner Refer to: Electrical Components Top Cover Refer to: Main Enclosure Components Pressure Vessel Refer to: Pressure Vessel Components Heating Element Refer to: Electrical Components Pressure Relief Valve Pulse (Door)Solenoid Refer to: Plumbing & Sensor Components Refer to: Electrical Components Steam Trap D
Return To Table Of Contents SECTION VI PARTS LIST M11 / M11D M11D Component / Page Cross-Reference Labels Trays / Rack Display PC Board Refer to: Labels & Decals Refer to: Racks, Trays, Cleaner Refer to: Electrical Components Top Cover Refer to: Main Enclosure Components Pressure Vessel Refer to: Pressure Vessel Components Heating Element Refer to: Electrical Components Pressure Relief Valve Refer to: Plumbing & Sensor Components Door Switch Refer to: Electrical Components Bellows Refer to: Plum
Return To Table Of Contents SECTION VI PARTS LIST M11 / M11D Main Enclosure Components 2 1 8 3 4 14 5 4 20 7 9 24 28 8 27 10 15 32 19 29 8 14 12 27 13 8 14 30 6 15 26 11 18 21 31 18 25 23 8 21 11 7 16 17 14 22 MA225405 Used on units with Serial Numbers: ES1347 thru ES3020 ET1095 thru ET3317 GC1000 thru GC1018 GD1000 thru GD1015 Item 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 Part No. Description 030-1117-00 050-2638-00 040-0008-87 040-0010-88 Top Inspection Cover ...............
Return To Table Of Contents SECTION VI PARTS LIST M11 / M11D Main Enclosure Components 2 1 8 3 4 14 5 4 20 7 9 24 28 8 27 10 15 32 19 29 8 14 12 27 13 8 14 30 6 15 26 11 25 21 31 18 18 23 8 21 11 7 16 17 14 22 MA225405 ES3021 thru ES3376 FR1000 thru FR1100 Item 1 2 3 4 5 6 7 8 9 10 11 12 13 14 Used on units with Serial Numbers: ET3318 thru ET3893 FP1000 thru FP1063 GB1000 thru GB1044 GC1019 thru GC1020 Part No.
Return To Table Of Contents SECTION VI PARTS LIST M11 / M11D Main Enclosure Components 2 1 8 3 4 14 5 4 20 7 8 24 28 27 10 15 31 19 9 29 8 14 12 27 13 8 14 6 15 26 11 25 21 30 18 18 23 8 21 11 7 16 17 14 22 MA225406 Used on units with Serial Numbers: ES3377 thru ES4879 GB1043 thru GB1109 Item 1 2 3 4 5 6 7 8 9 10 11 12 13 14 ET3319 thru ET6315 GC1020 thru GC1074 Part No. Description 030-1117-00 050-2638-00 040-0008-87 040-0010-88 Top Inspection Cover ....................
Return To Table Of Contents Main Enclosure Components SECTION VI PARTS LIST M11 / M11D 2 1 8 3 4 14 5 4 20 7 8 24 28 27 10 15 31 19 9 29 8 14 12 27 13 8 14 6 15 26 11 25 21 30 18 18 23 8 21 11 7 16 17 22 14 MA225408 Used on units with Serial Numbers: ES4880, Item 1 2 3 4 5 6 7 8 9 10 11 12 13 14 ET6316, FP6316, FR1204, GB1110, Part No. Description 030-1117-00 050-2638-00 040-0008-50 040-0010-88 Top Inspection Cover ............................
Return To Table Of Contents SECTION VI PARTS LIST M11 / M11D Plumbing & Sensor Components 21 25 20 19 18 28 33 27 24 10 28 15 29 19 16 22 9 8 26 23 17 4 3 2 1 A 6 C 30 31 32 34 11 B 12 13 7 5 14 MA201915 Used on units with Serial Numbers: ES1000 thru ES4300 GB1000 thru GB1094 Item 1 2 3 5 6 7 8 Part No.
Return To Table Of Contents SECTION VI PARTS LIST M11 / M11D Plumbing & Sensor Components 21 25 20 19 18 28 33 27 24 10 28 15 29 16 22 9 8 26 23 19 17 4 3 2 1 A 6 C 30 31 32 34 11 B 13 7 5 12 14 MA201915 Used on units with Serial Numbers: ES4301 thru ES5751 GB1095 thru GB1129 Item ET5400 thru ET7341 GC1036 thru GC1093 FP1073 thru FP1075 GD1050 thru GD1102 Part No.
Return To Table Of Contents SECTION VI PARTS LIST M11 / M11D Plumbing & Sensor Components 20 24 19 22 26 32 18 17 25 23 9 27 28 14 18 15 8 7 21 16 27 3 2 1 A 5 C 29 30 31 10 33 B 11 12 6 4 13 MA511004 Used on units with Serial Numbers: ES5752 thru ES8308 FR1147 thru FR1412 Item ET7342 thru ET12157 GB1129 thru GB1130 Part No.
Return To Table Of Contents SECTION VI PARTS LIST M11 / M11D Plumbing & Sensor Components 20 24 19 22 25 32 18 26 23 9 14 8 17 18 15 7 27 28 21 16 25 3 2 1 A 5 C 29 30 31 10 33 B 11 6 4 13 12 MA511005 Used on units with Serial Numbers: ES8309, Item ET12158, FP1149, FR1413, GB1131, Part No.
Return To Table Of Contents Pressure Vessel Components SECTION VI PARTS LIST M11 / M11D 21 22 20 24 2 23 3 38 26 1 4 30 36 5 3 27 28 6 29 7 40 11 9 33 10 40 14 34 13 15 39 31 12 25 8 16 17 37 35 32 18 WARNING Do not attempt to adjust, modify, or alter in any manner, any part of the pressure vessel. Serious injury and/or damage to the unit could result.
Return To Table Of Contents Pressure Vessel Components SECTION VI PARTS LIST M11 / M11D 21 22 20 24 2 23 3 26 1 4 30 36 5 3 27 28 6 29 7 11 40 9 33 10 39 14 34 13 15 38 31 12 25 16 17 37 8 35 32 WARNING 18 Do not attempt to adjust, modify, or alter in any manner, any part of the pressure vessel. Serious injury and/or damage to the unit could result.
Return To Table Of Contents SECTION VI PARTS LIST M11 / M11D M11 Manifold Components A 3 2 4 5 1 6 7 C 9 B 8 NOTE When replacing components in the solenoid, examine the existing valve to determine if it was Full Wave Rectified (if so will be marked FWR). MA202201 Used on Units with Serial Numbers ES1000 thru ES1346 ET1000 thru ET1094 Item Part No. Description Qty.
Return To Table Of Contents SECTION VI PARTS LIST M11 / M11D M11 Manifold Components A 3 4 5 2 7 6 C 8 9 B 1 NOTE When replacing components in the solenoid, examine the existing valve to determine if it was Full Wave Rectified (if so, will be marked FWR). MA202206 Used on units with Serial Numbers: ES1347 thru ES5751 FR1000 thru FR1246 Item Part No. Description 029-1271-00 1 2 3 4 ET1095 thru ET7341 GB1000 thru GB1129 Qty.
Return To Table Of Contents SECTION VI PARTS LIST M11 / M11D M11 Manifold Components A 4 3 5 2 7 6 C 8 B 1 NOTE When replacing components in the solenoid, examine the existing valve to determine if it was Full Wave Rectified (if so, will be marked FWR). MA511400 ES5752, Item Part No. Used on units with Serial Numbers: ET7342, FP1076, FR1247, GB1130, NP1000 - thru present Description Qty.
Return To Table Of Contents SECTION VI PARTS LIST M11 / M11D M11D Manifold Components A 1 4 5 6 2 3 8 7 C 10 9 B 11 NOTE When replacing components in the solenoid, examine the existing valve to determine if it was Full Wave Rectified (if so, will be marked FWR). MA202207 Used on Units with Serial Numbers GC1000 thru GC1093 GD1000 thru GD1102 Item 1 2 3 4 5 6 7 Part No. 014-0183-01 002-0519-00 • 014-0275-02 014-0194-00 014-0276-00 014-0234-00 014-0184-01 Description Qty. Male Connector ......
Return To Table Of Contents SECTION VI PARTS LIST M11 / M11D M11D Manifold Components A 1 4 5 6 2 3 8 7 C 10 9 B 11 NOTE When replacing components in the solenoid, examine the existing valve to determine if it was Full Wave Rectified (if so, will be marked FWR). MA510900 Used on Units with Serial Numbers GC1094 & GD1103 - thru present Item 1 2 3 4 5 6 7 Part No. 014-0183-01 002-0519-00 • 014-0275-02 014-0194-00 014-0276-00 014-0234-00 014-0184-01 Description Qty. Male Connector ...............
Return To Table Of Contents SECTION VI PARTS LIST M11 / M11D Electrical Components 9 1 33 13 8 6 2 10 12 19 7 18 14 11 15 22 23 24 31 32 21 17 16 20 26 25 30 5 27 28 3 4 29 MA202305 Used on Units with Serial Numbers ES1000 thru ES1346 ET1000 thru ET1094 Item 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 Part No.
Return To Table Of Contents SECTION VI PARTS LIST M11 / M11D Electrical Components 10 9 8 1 6 13 7 2 11 12 19 14 15 22 23 24 18 31 32 17 21 16 20 26 25 30 5 27 28 3 4 29 MA202308 Used on units with Serial Numbers: ES1347 thru ES3020 Item 1 2 3 4 5 6 7 8 9 10 Part No.
Return To Table Of Contents SECTION VI PARTS LIST M11 / M11D Electrical Components 10 9 8 1 6 11 7 33 2 12 18 13 21 17 22 30 31 23 14 16 20 15 19 32 25 1 24 29 5 26 27 3 4 28 MA374706 Used on units with Serial Numbers: ES3021 thru ES7392 FR1000 thru FR1356 Item 1 2 3 4 5 6 7 8 9 10 Part No.
Return To Table Of Contents SECTION VI PARTS LIST M11 / M11D Electrical Components 10 9 2 8 6 30 11 7 1 18 12 13 21 22 23 19 17 5 16 20 25 14 15 24 29 26 27 3 4 28 MA374707 Used on Units with Serial Numbers: GC1019 thru GC1155 GD1016 thru GD1192 Item Part No. 1 2 3 4 5 6 7 8 9 10 11 015-1137-00 002-0370-00 057-0426-00 041-0250-18 040-0010-75 040-0006-04 002-0369-02 053-0454-00 061-0531-00 12 015-1096-00 13 14 15 040-0010-46 015-0657-00 Description Qty.
Return To Table Of Contents SECTION VI PARTS LIST M11 / M11D Electrical Components 10 9 8 1 6 11 7 31 2 12 18 13 21 14 17 22 16 23 20 15 19 30 25 1 24 29 5 26 27 3 4 28 MA374708 Used on units with Serial Numbers: ES7393, Item 1 2 3 4 5 6 7 8 9 10 Part No.
Return To Table Of Contents SECTION VI PARTS LIST M11 / M11D M11 Control PC Board 1 2 4 3 MA222001 Used on Units with Serial Numbers ES1000 thru ES1346 ET1000 thru ET1094 Item 1 Part No. Description 002-0501-00 Control PC Board - 115 V (Includes Items 1thru 3) {Old Board Replaced By New Board) ..................... 1 • M11 EPROM .................................... Ref. • Qty. Item Part No. Description Qty. 2 3 • 015-0346-08 • 015-0346-07 4 • • Printer Fuse (3 amp/250 V) .............
Return To Table Of Contents SECTION VI PARTS LIST M11 / M11D M11 Control PC Board 1 2 3 MA281100 Used on Units with Serial Numbers: GB1000 thru GB1109 Item 1 2 Part No. Description 002-0501-01 M11 Control PC Board {230 V Domestic} (Includes Item 1thru 3) .......................... 1 • M11 EPROM .................................... Ref. • Printer Fuse (3 amp/250 V) ................. 1 • • 015-0346-08 Qty. Item 3 Part No.
Return To Table Of Contents SECTION VI PARTS LIST M11 / M11D M11 Control PC Board 1 2 3 MA281100 Used on Units with Serial Numbers: ES1347, ET1095, FP1000, FR1000, GB1110, NP1000 - thru present Item Part No. Description 002-0501-00 M11 Control PC Board {115 V Domestic} (Includes Item 1thru 3)[Used on Serial Number Prefix "ES & ET"] ........... 1 M11 Control PC Board {230 V Domestic} (Includes Item 1thru 3) [Used on Serial Number Prefix "FP & FR"] ...........
Return To Table Of Contents SECTION VI PARTS LIST M11 / M11D M11D Control PC Board 1 2 3 Item 1 2 Part No. Description 002-0501-02 M11D Control PC Board {115 V Domestic} (Includes Item 1thru 3) .......................... 1 • M11 EPROM .................................... Ref. • Printer Fuse (3 amp/250 V) ................. 1 • • 015-0346-08 Qty. Item 3 Part No. Description • 015-0346-07 • Line Fuse (15 amp/250 V) {Used on 115 V Units Only} ................................
Return To Table Of Contents SECTION VI PARTS LIST M11 / M11D M11 Labels and Decals 5 12 11 4 6 9 2 10 7 13 1 3 8 MA202512 Item 1 2 3 4 5 6 7 Part No. Description 061-0203-00 061-0516-00 061-0477-00 Electrical Shock Label (English) ............ Electrical Shock Label (French) ........... Patent Label .......................................... Serial Number Label (Front) .................. Serial Number Label (Rear) ................... M11 Warning Label P.R. Valve(English) M11 Warning Label P.R.
Return To Table Of Contents SECTION VI PARTS LIST M11 / M11D M11D Labels and Decals 5 11 9 6 4 10 2 7 1 3 8 MA202511 Item 1 2 3 4 5 6 7 Part No. Description 061-0203-00 061-0477-00 Electrical Shock Label ........................... Patent Label .......................................... Serial Number Label (Front) .................. Serial Number Label (Rear) ................... Warning Label P.R. Valve ...................... Warning Label .......................................
Return To Table Of Contents SECTION VI PARTS LIST M11 / M11D Printer Components 3 4 2 5 8 7 OF F ON /F EE D 1 MA203302 Item 1 2 3 4 5 Part No. Description 9A155001 • 053-0506-00 • 060-0008-00 • 053-0507-00 • 053-0505-00 Accessory Printer (Incl. Items 2 thru 7) . • Arbor ................................................... • Paper Roll ........................................... • Cover .................................................. • Printer Cartridge ..................................
Return To Table Of Contents SECTION VI PARTS LIST M11 / M11D Racks, Trays, and Cleaner MA203205 Used on units with Serial Numbers: ES1000 thru ES5001 GB1000 thru GB1109 Item 1 2 3 4 5 6 ET1000 thru ET6434 GC1000 thru GC1074 Part No. Description 9A215001 9A215002 050-3920-00 050-3919-00 050-2644-00 002-0269-00 M11 Horizontal Cassette Rack Acces. .. 1 M11 Vertical Cassette Rack Acces. ...... 1 Wire Tray Rack ..................................... 1 Tray Plate ...........................................
Return To Table Of Contents SECTION VI PARTS LIST M11 / M11D Racks, Trays, and Cleaner MA203205 Used on units with Serial Numbers: ES5002, Item 1 2 3 4 5 6 ET6435, FP1074, FR1205, Part No. Description 9A215001 9A215002 050-3920-00 050-3919-00 050-4260-00 002-0269-00 M11 Horizontal Cassette Rack Acces. .. 1 M11 Vertical Cassette Rack Acces. ...... 1 Wire Tray Rack ..................................... 1 Tray Plate ............................................. 1 Small Tray ..........................
Return To Table Of Contents SECTION VI PARTS LIST M11 / M11D Packaging 7 5 6 4 3 1 2 Item 1 2 3 4 5 6 7 8a MA202405 Part No. Description Qty. 003-1030-00 003-1032-00 063-9983-00 066-0844-00 054-0317-00 053-0061-01 077-0033-00 054-0317-00 061-0719-00 M11 Installation & Operation Manual ..... M11D Installation & Operation Manual .... Skid ........................................................ Cardboard Box ...................................... Bottom Pad (includes item 7) .................
Return To Table Of Contents SECTION VI PARTS LIST This page intentionally left blank.
Return To Table Of Contents COMMENTS The Technical Publications Department of Midmark Corporation takes pride in its publications. We are sure that our manuals will fill all of your needs when you are performing scheduled maintenance, servicing, or repairs on a Midmark product.
Return To Table Of Contents FAX ORDERING FORM (SERVICE PARTS ONLY) NOTES: • • ALL BLOCKED AREAS MUST BE COMPLETED. USE FOR NON-WARRANTY FAX ORDERS ONLY. WARRANTY ORDERS MUST BE TELEPHONED IN (1-800-MIDMARK). ATTENTION: SERVICE DEPARTMENT FAX#: 877-249-1793 ACCT #: P.O. #: DATE: NAME: SHIP TO: ADDRESS: CITY, ST.: CONTACT: PHONE: NON-EMERGENCY ORDER - TO SHIP WITHIN 72 HOURS IF PART(S) IN STOCK. EMERGENCY ORDER - TO SHIP WITHIN 24 HOURS IF PART(S) IN STOCK (IF ORDER IS RECEIVED BEFORE 1:00 P.M. E.S.
Return To Table Of Contents
Return To Table Of Contents Midmark Corporation 60 Vista Drive P.O. Box 286 Versailles, Ohio 45380-0286 937-526-3662 Fax 937-526-5542 midmark.com Subject to change without notice. Refer to www.Documark.com for latest revision. © Midmark Corporation - 2005 Printed in U.S.A.