Specifications

25
Sterilization failure
evidence from process
monitor (chemical
indicator, biological
indicator, etc.)
sterilization conditions
were not present at the
location of the indicator.
reload the sterilizer in
accordance with Guidelines
for Loading Trays. If the
problem recurs, take the
unit out of service and
contact an authorized
service representative (see
Calling For Service later in
this manual).
insufficient air removal, low
temperature, or low
pressure.
take the unit out of service
and contact an authorized
service representative (see
Calling For Service later in
this manual).
indicator is out of date, is
inappropriate for sterilizer
cycle, or has
malfunctioned.
use an indicator,
appropriate for the load and
cycle selected, from a fresh
supply that has been stored
properly. Contact the
indicator manufacturer for
additional information on
proper selection, use,
storage, and potential
misapplication or
malfunction.
Door gasket leaks. door gasket is damaged or
dirty.
allow sterilizer to cool; then
clean or replace door
gasket (see Remove and
Clean Door Gasket under
Quarterly maintenance in
this manual).
sterilizer is not level. level the sterilizer.
Packs not dry sterilizer is overloaded. reload the sterilizer in
accordance with Guidelines
for Loading Trays. If the
problem recurs, take the
unit out of service and
contact an authorized
service representative (see
Calling For Service later in
this manual).
chamber filter is clogged. allow sterilizer to cool; then
clean or replace filter (see
Cleaning Chamber Filter
under Monthly maintenance
in this manual).
Troubleshooting Guide - (Continued)
Problem Possible Cause Solution
Operator
Maintenance
Return To Table Of Contents