Specifications
24
(a)Open sterilizer door. Then,
remove dam gasket (1) and
door gasket (2) from
door(3).
(b)Clean the gaskets (1 and 2)
with a mild detergent and
inspect the gaskets for
cracks, cuts, shrinkage, or
swelling. Replace gaskets if
necessary.
(c)Clean the gasket housing
channel (A) with a solution
of mild soap and distilled or
demineralized water. A
small brush (not wire brush)
may be used to aid the
cleaning.
(d)Press the door gasket (2)
into the channel (A). Then,
install dam gasket(1).
(e)Run one cycle to seat the gaskets properly.
Troubleshooting Guide
Use the following table to assist in correcting minor problems with the sterilizer.
Troubleshooting Guide
Problem Possible Cause Solution
WATER LOW lamp in
on.
not enough water in
reservoir tank to fill
chamber.
add distilled or
demineralized water to
reservoir tank.
DOOR AJAR lamp is on. Chamber door is not fully
closed and / or latched.
Close the door properly and
make sure the door is fully
latched.
ERROR lamp is on. _ See Error Code Guide for a
detailed explanation
Steam is escaping from
the pressure relief valve.
pressure relief valve was
not properly reseated after
pressure relief valve
inspection
Reseat the pressure relief
valve (see Perform Pressure
Relief Valve Check under
Monthly maintenance in this
manual).
3
2
1
A
MA438700
Operator
Maintenance