Error Messages - continued
Calling for Service
Contact your Midmark Authorized Dealer,orlogontowww.midmark.com.
Model and serial number information will be required whencallingforservice.
To contact Midmark directly:
1-800-Midmark (1-800-643-6275) or 937-526-3662
8:00amto5:00pmEST(MondaythruFriday)
[excluding standard U.S. holidays]
Note
Please mark down any displayed code(s) and be sure to relay this information to the service technician.
Message Possible Cause Solution
C642 THROUGH C647
PRESSURE HIGH
Chamber pressure is
outside the limits for
normal operation.
Waitbriey(uptooneminute)whilechamberpressure
/ temperature dissipates. Press STOP button to return
to select cycle mode. Initiate a new cycle.
C660 THROUGH C677
PRESSURE HARDWARE
or
PRESSURE OVERLIMIT
Chamber pressure is
outside the limits for
normal operation.
Unplugunitpowercordfor1minuteandthenplug
backin.Ifproblempersists,contactanauthorized
servicerepresentative.(see“CallingForService”)
C980 THROUGH C987
HI-LIMIT OPEN
Highlimitswitchhas
opened for at least 1/4
secondduringspecic
operational mode.
Unplugunitpowercordfor30minuteandthenplug
backin.Ifproblempersists,contactanauthorized
servicerepresentative.(see“CallingForService”)