Specifications

42
TP202 Rev. A © Midmark Corporation 2013
Error Messages - continued
Message Possible Cause Solution
CO60: POWER UP MODE
SYSTEM HARDWARE
Power was interrupted
brieyoraninternal
glitch.
Unplugunitpowercordfor1minuteandthenplug
backin.Ifproblempersists,contactanauthorized
service
representative(see“CallingForService”).
C102: FILL MODE
STOP PRESSED
STOP button was pressed
duringcycle.
Press STOP button to restart.
C103: THROUGH C105
HEATUP, STERILIZE OR
VENT MODE
STOP PRESSED
STOP button was pressed
duringcycle.
Waitbriey(uptooneminute)whilechamberpressure
/ temperature dissipates. Press STOP button to return
to select cycle mode where a new cycle may be initi-
ated.
C106: DOOR MODE
STOP PRESSED
STOP button was pressed
duringcycle.
Press STOP button to restart.
C232: FILL MODE
WATER LOW
Waterlevelin
reservoir is too low.
Rellwaterreservoirwithdistilledordemineralized
water.Waitbriey(uptooneminute).PressSTOP
button to return to select cycle mode where a new
cycle may be initiated.
Fill/Ventlter(inbottom
ofchamber)isclogged.
CleanFill/Ventlter.
(see“Non-ScheduledMaintenance”)
C326: DOOR MODE DOOR
CLOSED
Door latch safety switch is
stillmakingcontactafter
door motor operated.
Open door.
C382: FILL
DOOR OPEN
Sterilizer detects that door
safety switch contacts
opened.
Close the sterilizer door.
(Cycle will continue where left off.)
C383: HEATUP MODE
DOOR OPEN
Sterilizer detects that door
safety switch contacts
opened.
Waitbriey(1minute)whilechamberpressure/tem-
perature dissipates. Press STOP button to return to
select cycle mode. Initiate a new cycle.
C384: STERILIZE MODE
DOOR OPEN
Sterilizer detects that door
safety switch contacts
opened.
Unplugunitpowercordfor1minuteandthenplug
back in.
Ifproblempersists,contactanauthorizedservice
representative(see“CallingForService”).
C533 through C633:
STEAM TEMP LOW
or
STEAM TEMP HARDWARE
or
PRESSURE LOW
Sterilizer detects that
temperature and / or
pressure is outside the
limits for normal
operation.
Unplugunitpowercordfor1minuteandthenplug
backin.Ifproblempersists,contactanauthorized
servicerepresentative(see“CallingForService”).
This chart continues on the following page...