Specifications
41
TP202 Rev. A © Midmark Corporation 2013
Informational Messages
Thechartbelowliststheinformationalmessagesthatmayappearduringoperation.
Error Messages
Thechartbelowandonthefollowingpages,liststheerrormessagesthatmayappearduringoperation.
Message Possible Cause Solution
INTIALIZING SYSTEM
Unitpowercordwasjustplugged
in - standard informational
message.
Normal operation will occur after a
4 second pause.
TOTAL CYCLES
XXX M9*, VX.X.X
(*will display model
number,M9,M11)
Unitpowercordwasjustplugged
in - standard informational
message.
Normal operation will occur after this
message.
PERFORM WEEKLY
MAINTENANCE
Thismessageisdisplayedevery
7,14,and21daysaftertheunit
ispluggedintoapowersourceto
prompt the operator to preform
weekly maintenance described in
this manuals
Preform weekly maintenance.
Themessagewillautomaticallyclear
afterthenextcycleiscompleted.
PERFORM MONTHLY
MAINTENANCE
Thismessageisdisplacedevery
28 days to prompt operator to
perform monthly maintenance
described in this manual.
Perform monthly maintenance.
Themessagewillautomaticallyclear
afterthenextcycleiscompleted
Caution
If an error occurs more than once, stop using the sterilizer. Record the message or error
code, unplug unit, then call an authorized service representative. (see “Calling for Service”)
If an error message includes: “Items Not Sterile”, the items in sterilizer shall not be considered sterile;
they must be run through a successful sterilization cycle.
Message Possible
Cause
Solution
C010: POWER UP MODE
SYSTEM PWR LOSS
ITEMS NOT STERILE
PUSH STOP TO RESTART
Unit had loss of power
duringcycle.
Press STOP button to restart.
This chart continues on the following page...