Specifications
40
TP202 Rev. A © Midmark Corporation 2013
Troubleshooting
Troubleshooting Chart
Usethefollowingcharttoassistincorrectingminorproblemswiththesterilizer.
Problem Possible Cause Solution
Sterilizer does not
operate (no display).
Sterilizer power cord came loose from
supply outlet or back of sterilizer.
AssureSterilizerpowercordispluggedinto
outlet and sterilizer.
No power at Sterilizer supply outlet. Check circuit breaker for supply outlet. If
problemrecurs,unplugunitpowercordand
contact an authorized service technician.
(see“CallingforService”).
Fuse open on main P.C. Board. Unplugunitpowercordandcontactan
authorized service technician
(see“CallingforService”).
Steam escaping
from pressure relief
valve.
Pressure relief valve not properly seated. Reseat pressure relief valve
(see“MonthlyMaintenance”).
Sterilization
failure evidence
from process
monitor (chemical
indicator, biological
indicator, etc.).
Sterilization conditions were not present
at location of the indicator.
Reloadsterilizerinaccordancewithguidelines
forloadingtheM9orM11.Ifproblemsrecurs,
unplugunitpowercordandcontactanautho-
rized service technician
(see“CallingforService”).
Indicatorisout-of-date,isinappropriate
for sterilization cycle. or has otherwise
malfunctioned.
Useanindicator,appropriatefortheloadand
cycleselected,thathasbeenstoredproperly.
Contact the indicator manufacturer for additional
informationonproperselection,use,storage,
and potential misapplication or malfunction.
Door gasket leaks.
Doorgasketisdamagedordirty. Cleanorreplacedoorgasket
(see“WeeklyMaintenance”).
Items are not dry at
end of Dry Cycle.
Sterilizer is improperly loaded. Reloadsterilizerinaccordancewithguidelines
forloadingtheM9orM11.
Ifproblemsrecurs,unplugunitpowercordand
contact an authorized service technician
(seecallingforservice).
Filterscreen(s)clogged.(Checkll/Vent
andAirValvelterscreens)
Cleanorreplacelterscreen(s).
(see“Non-ScheduledMaintenance”)
Printer not printing.
Printer cable is not connected properly. Ensurethatprintercableisconnectedtoprinter
and PC board properly.
Software malfunction. Unplugsterilizerpowercord,wait15seconds,
andthenplugsterilizerpowercordbackin.
Printer is out of paper. Insertanewpaperroll(seeinsertingthepaper
roll under printer installation and operation)