Specifications
36
TP202 Rev. A © Midmark Corporation 2013
Example of Typical Printout of a Program Cycle
Midmark M9 - v1.0.3
Total Cycles: 10
Sterilizer ID
Operator
mm / dd / yyyy hh : mm
BEGINUNWRAPPEDCYCLE
Temp: 270 DegreesF
Time: 3 Minutes
Dry: 30 Minutes
FILLINGCHAMBER
HEATING
mm:ss Degrees PSI
0:00 233.8 F 0.0
0:30 231.3 F 1.9
2:30 265.7 F 25.0
3:00 271.3 F 28.8
STERILIZING
mm:ss Degrees PSI
0:00 271.8 F 28.9
0:30 272.4 29.5
2:30 272.7 29.3
3:00 272.6 29.3
Min 271.8 F 28.7
Max 273.2 29.9
VENTINGCHAMBER
DRYING
MM:SS
0:00 DryingStart
5:00
25:00
30:00 DryingComplete
FILLING: 3:22
HEATING: 8:54
STERILIZING: 3:00
VENTING: 2:16
DRYING: 30:00
TOTALCYCLE 00:47:32
CYCLECOMPETE
- - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - -
- - / - - / - - - - - - : - -
Line provided for
sterilizer identication
Total # of cycles
run on unit
Line provided for
operator signature
Space provided for
recording date & time
Selected Cycle
Indicates Filling
phase initiated
Sterilization temperature,
time, and dry time settings
During Heat-Up phase, the
printer records chamber
temperature & pressure in
30-second increments
During Sterilization phase,
the printer records chamber
temperature & pressure in
30-second increments
Summary of
Sterilization phase
Indicates Venting
phase initiated
During Drying phase,
the printer records time in
5-minute increments
These lines will show the
total dry time completed
These lines will not print
if Dry Cycle is aborted
before it completes
Cycle summary