Specifications

35
TP202 Rev. A © Midmark Corporation 2013
Printer (Optional)
Operating the Printer
Whenthesterilizerispluggedin,theprinteris automatically powered up and initialized.
No user intervention or setup is required.
Power-Up Message
Whentheprintersuccessfullyinitializesafterthesterilizerispoweredup,itprintstheword“READY”to
assuretheoperatorthatitsbuiltinmicroprocessorisworkingproperlyandthesterilizationcycleswillbe
recorded by the printer.
Printer Tape Description
Theprinterwillprintthefollowinginformationforeachprogramcycle:
•CycleNumber
•Identication:alineisprintedsothesterilizerI.D.canberecordedontheprintertape
• Operator:alineisprintedsotheoperator’ssignaturecanberecordedontheprintertape.
•Date:linesareprintedforthemonth,day,year,hour,andminutesothedateandtimecanbe
recorded by the operator on the printer tape.
•“BEGINselectedCYCLE”:toindicatedthebeginningofthecycleselectedbytheoperator.
•Summaryofselectedcyclesetpoints.
Oncethecyclestarts,theprinterwillprintthewords“FILLINGCHAMBER”toshowthatthesterilizerisling
with water.
OncethesterilizerbeginstheHeatingPhaseofthesterilizationcycle,theword“HEATING”isprintedand
theprinterwillprintthechambertemperature,pressure,andelapsedtimein30sec.incrementsuntilthe
heatingphaseiscompleted.
Whenthesterilizerentersthesterilizationphase,theword“STERILIZING”isprintedandtheprinterwillprint
thechambertemperature,pressure,andelapsedtimein30sec.incrementsuntilthesterilizationphaseis
completed.
Whenthesterilizerhascompletedthesterilizationphaseoftheprogramcycle,theprinterwillprintthewords
“VENTINGCHAMBER”toshowthatthesteampressureisbeingexhaustedfromthechamber.
Whenthesterilizerentersthedryingphase,theword“DRYING”isprintedandtheprinterwillprintthe
elapsedtimein5minuteincrementsstartingwith0:00andthewords“DRYINGSTART”.Theprinter
continuestoprinttheelapsedtimein5minuteincrementsuntilthedryingphaseiscompleted.Thenal
recordforthedryingphasewillincludethewords“DRYINGCOMPLETE”.Intheeventthedryingtimeis
programmedtoatimethatisn’tdivisibleby5,thenalprintedrecordforthedryingphasewillreectthe
actualprogrammeddryingtimein1minuteincrements,e.g.aprogrammeddrytimeof12minuteswillhave
5,10,and12minutesprintedontheprintertape.
Whenthesterilizerhascompletedthedryingphaseofthesterilizationcycle,theprinterwillprintasummary
ofthesterilizationcyclewiththedurationofeachphaseofthecycleandtheTotalCycleTime.Followingthe
summarytheprinterwillprint“CYCLECOMPLETE”.
[NOTE: If drying cycle is aborted, “DRYING COMPLETE” and “CYCLE COMPLETE” will not print].