Specifications
33
TP202 Rev. A © Midmark Corporation 2013
Monthly Maintenance - continued
• Remove / Clean Door & Dam Gaskets
A) Remove door and dam gaskets from chamber door,
then remove the gasket ring from the door gasket.
B) Clean gaskets and ring with SpeedClean,
distilled water, and a soft brush.
C) Inspect gaskets for damage / shrinking / swelling.
Replace gaskets if damage is apparent.
D) Press gasket ring into the channel in the
door gasket and reinstall the gasket in the door.
E) Install dam gasket.
J) Grasp filter and gently pull away from chamber wall while twisting slightly.
(If necessary, pliers may be used to remove filters)
K) Clean filters with SpeedClean and distilled water. A small stiff bristled brush or
ultrasonic cleaner may be helpful. Rinse filters with distilled water.
(Note: Replace filter(s) if debris cannot be remvoved by cleaning).
L) Wipe out the inside of chamber.
M) Install filters. (Press inward, toward chamber wall while twisting slightly).
N) Install tray plate, rack, and trays.
Equipment Alert
Use care when wiping the inside of the chamber. Failure to comply may result in
damage to the heating element, steam temperature probe, and/or level sensor probe.
Equipment Alert
Do not operate sterilizer without filters in place.