Specifications
32
TP202 Rev. A © Midmark Corporation 2013
MA8195-1i
Monthly Maintenance
• Clean Chamber / Plumbing
A) With a cooled chamber, drain the sterilizer’s reservoir and refill with clean, distilled water.
Add one ounce of SpeedClean sterilizer cleaner directly to the bottom of chamber.
B) Run one Pouches cycle.
C) Press Stop button when Dry Cycle begins.
(It is not necessary for the dry cycle to run during this maintenance.)
D) Drain reservoir and refill a second time with clean, distilled water for the rinse cycle.
E) Rinse by running one Unwrapped cycle
, but push the “Stop” button when the drying cycle
begins.
F) Drain and refill reservoir with clean distilled water, then allow sterilizer to cool.
G) Remove trays and tray rack. Wipe off with a damp cloth.
H) Remove and clean filters. The filters are intended to prevent debris from causing valve failures.
Between regular monthly cleanings if the fill or vent times become too long or items will not dry the
filters should be cleaned.
I) Refer to the illustration for location of filter screens.
WARNING
Do not process instruments while flushing system.
Note
Every 28 days, the sterilizer will automatically display the PERFORM MONTHLY MAINTENANCE
message. This is a more thorough cleaning of the unit and involves multiple steps. If power is
disconnected, the cycle of monthly messages will be reset.
Fill / Vent
Filter Screen
Steam Temperature Probe
Level Sensor Probe
Heating Element
SpeedClean
sterilizer cleaner
Air Filter Screen