Specifications
30
TP202 Rev. A © Midmark Corporation 2013
Weekly Maintenance
• Clean Chamber / Trays (including Rack and Plate)
A) Disconnect the upper portion of the reservoir drain tube from the panel clips, bend it downward,
and drain the reservoir water into a suitable container, e.g. a bucket, and dispose of the water.
B) Remove the trays, tray rack, and tray plate from the sterilizer.
(Refer to the following page for instructions on removing / installing the tray rack and tray plate).
C) Wash trays, rack, plate, and inside of chamber with mild soap or Speed-Clean and distilled water.
D) Refill reservoir with distilled water.
Equipment Alert
Failure to change water may result in sterilizer malfunction.
Do not use bleaching agents or any abrasive materials / substances in chamber
(i.e. bleach, steel wool, wire brush, scouring powder, etc.).
Failure to comply may result in damage to the chamber and/or other components.
Tray Rack
Tray Plate
Trays
Note
Everysevendays,theautoclavewillautomaticallydisplaythePERFORMWEEKLYMAINTENANCE
message.Ifpowerisdisconnected,thecycleofweeklymessageswillbereset.