Specifications

28
TP202 Rev. A © Midmark Corporation 2013
NOTE:
PressingtheSTOPbuttonduringthisprocedurewillabortthechanges,andreverttotheoriginalsettings.
Action Description Display
Adjust Dry Time Dry time can be adjusted from 0 to 60
minutes.
The“+”and“-”buttonsadjustthe
time by 1 minute increments.
Whenthedesiredtimeappearsonthe
display...
Press Tostorethesettingandcompletethepro-
grammingprocess.Thedisplaywillshow
the new cycle parameters.
Note
The programmed settings are retained under Program # button <1> or <2>.
Even if power is interrupted, or the unit is unplugged the setting will be retained.
Caution
All materials processed using these cycles must be validated by the user.
Failure to do so could result in incomplete sterilization.
Programmable Cycle Buttons - continued
Maintenance
Maintenance Messages
Toassureproperoperationandmaximumsterilizerlife,carefullyfollowallrecommendationsforperiodic
maintenance. Recommended maintenance is easy to do and takes very little time.
One of the MOST important steps you can take to prevent problems with your sterilizer is to ensure that
ONLYdistilledwater-NOTTAPWATER-isusedinthesterilizer.Sincethesterilizeroperateswithhigh
watertemperatures,anymineralsdissolvedinthewaterwillformmineraldeposits.Thiscanpreventvalves
fromopeningorclosingproperlyandcanalsoleadtocorrosioninthechamberandtubing.
Maintenance reminders will be displayed on the screen at the appropriate intervals to assist the operator.
These reminders are removed from the display screen once a cycle is started.
Theuserisresponsibleforestablishingaperiodicmaintenanceproceduretoassurecorrectoperationof
equipment and reliable sterilization of loads. Contact your local Midmark distributor or service representative
todevelopaprogramforplannedmaintenance.