Specifications
27
TP202 Rev. A © Midmark Corporation 2013
NOTE:
PressingtheSTOPbuttonduringthisprocedurewillabortthechanges,andreverttotheoriginalsettings.
Action Description Display
Press Tostorethetemperature.Thisbringsup
thesterilizationtimeprogramingdisplay
showingthecurrentvalue.
Caution
Sterilization time can be adjusted from a minimum of 3 minutes to a maximum
of 90 minutes. It is important to properly coordinate the cycle time with the
sterilization temperature.
Adjust Sterilization
Time
The“+”and“-”buttonsadjustthe
time by 1 minute increments.
Whenthedesiredtimeappearsonthe
display...
Press Tostorethesterilizationtime.Thisbrings
uptheventspeedprogramingdisplayand
the current value.
Adjust Vent Speed InFASTvent,thevalvefullyopensand
vents the chamber.
InSLOWvent,thevalveopensfora
fraction of a second (once per minute)
to slowly vent the chamber.
Pressing“+”setsitto:FAST
Pressing“-”setsitto:SLOW
Whenthedesiredsettingappearsonthe
display...
Press Tostorethesettingandbringupthedry
timeprogramingdisplayandthecurrent
value.
This chart continues on the following page...
Programmable Cycle Buttons - continued