Specifications
22
TP202 Rev. A © Midmark Corporation 2013
Cycle Operation - continued
Afterthe‘Start’buttonispressed,thestage/statusofthecurrentcyclewillappearonthedisplay.
Thechartbelowillustratesthedisplaymessagesthatwillappearduringeachstageofthecycle.
Stage of Cycle Description Display
Filling Chamberbeginsllingwithwater.
Whenwaterreachestheproperlevel...
Heating Displaychangesastemperatureand
pressureinchamberchanges.
(Metric units may be displayed if desired).
Sterilizing Sterilizingbeginswhencorrect
temperature / pressure is reached.
Timeremainingincyclecountsdown
while current temperature / pressure in
chamber is continuously updated.
Ready to Vent ‘READY TO VENT’ is displayed when
10 seconds remain in sterilization cycle.
Fast Vent Whentimerunsoutinsterilizingmode,
the vent valve opens. Steam / water are
released back into the reservoir.
Thedisplaychangesastemperatureand
pressureinchamberchanges.
Caution
Keep clear when M9/M11 door is ready to open!
Failure to do so could result in severe burns from steam being released.
Door To Open
(automatic)
Pertains only to
M9/M11 UltraClave™
Anaudiblesignalisemittedtoindicate
that the door is about to open.
Whenpressureinchamberreacheszero,
thedooractuatestopartiallyopen(drying
mode) position.
This chart continues on the following page...