Specifications

20
TP202 Rev. A © Midmark Corporation 2013
Cycle Operation
Refertothefollowingstepsforadetaileddescriptionofcycleoperation:
Note
Pressing the Start button when ‘SELECT CYCLE’ is
displayed, at beginning or end of a cycle, activates
the heater for 10 minutes. The display flashes
‘ADDITIONAL HEAT’. This allows the operator to
preheat the chamber before starting a cycle or to
add additional time to the Dry mode at the end of a
cycle. Pressing the Stop button will end the
‘ADDITIONAL HEAT’ time.
Equipment Alert
The sterilizer will not operate unless the door is closed and latched properly.
Step 1: Close and latch the door.
A) Lift the door handle, then push the door closed.
B) While pushing in on the door, slide the door handle down to engage the latch.
Equipment Alert
Using an incorrect sterilization program could result in non-sterile goods and may
damage instruments. Consult instrument manufacturer for specific sterilization instructions.
Step 2: Select the desired cycle.
A) Press the appropriate cycle button on the display panel.
(Refer to “Cycle Parameters” chart for time / temperature specifications).
Cycle Buttons
(Pre-set)
Cycle Buttons
(Programmable)