Specifications

13
TP202 Rev. A © Midmark Corporation 2013
Display:
WARNING
Check the serial number label on back panel of sterilizer to verify voltage rating for
the unit. Failure to connect sterilizer to an appropriate power supply could result in
damage to the unit, and electrical shock to personnel.
Voltage Rating
Connecting the Power Cord
Equipment Alert
For optimal performance, allow sterilizer to reach
room temperature before operating.
To connect the power cord...
A) Plug power cord into receptacle on back of sterilizer.
B) Plug power cord into a properly polarized and grounded receptacle rated for a
minimum of 15 amps. A dedicated circuit only used for the sterilizer is recommended.
Note: When power is connected, the messages shown below will appear on the display.
WARNING
Equipment is not suitable for use in the presence of a flammable anesthetic mixture
with oxygen, air, or nitrous oxide.
Clarification: Equipment is suitable for use in the presence of oxygen, air, or nitrous oxide.
*
* This screen will display the total number of cycles run on the unit,
the model number (M9 or M11), and the software version number.