Specifications

12
TP202 Rev. A © Midmark Corporation 2013
Support Surface
•Materialshouldbewater-resistantmaterial.(Ex.laminate,stainlesssteel,stone,etc.)
•Surfacemustbeleveltoensurechamberllswithcorrectwaterlevel.
Improper water level in the chamber could cause a sterilizer malfunction.
•Surfaceshouldmeetminimumdimensionslistedbelow:
Dimensions
Depth (front to back) M11 - 21” (53 cm) M9 - 18” (46 cm)
Clearance Requirements
Toensureproperaircirculation,andtoallowaccesstothereservoirllportanddraincoupling,
adhere to the minimum clearance requirements listed below. If the sterilizer will be operated
incontinuouscycles,locatesterilizerwheresteamwillnotdamagematerialsorequipmentin
thesurroundingarea.
BackofUnit-BackWall ................................. 2” (5 cm)
Front Support Surface - Front Sterilizer ........ 1” (3 cm)
SidesofUnit-SideWall ................................ 2” (5 cm)
Distance above Unit for Printer Access .......... 5” (13 cm)
MaximumUpperCabinetShelfOverhang ..... M11 - 18” (46 cm) M9 - 15” (38 cm)
Under Cabinet or Shelf................................... M11 - 23” (58 cm) M9 - 22” (56 cm)
Relocation Requirements for Sterilizer
Disconnect power cord from electrical outlet and allow sterilizer to cool.
Drainwaterfromreservoirordonottipsterilizer,allowingwatertospill.
Installation
Location Requirements continued...
WARNING
For 115 VAC models: Use 104 - 127 VAC, 50/60 Hz alternating current only.
For 230 VAC models: Use 207 - 253 VAC, 50/60 Hz alternating current only.
Failure to do so may result in electric shock to personnel and / or damage to sterilizer.
Note
For safety, the unit must be connected to a properly polarized and grounded receptacle. Always use a
power cord with grounding connections that match the receptacles in your location.
115 VAC Unit: 115VAC,50/60Hz,12amp
DedicatedSupplyCircuit*:120VAC,50/60Hz,15amp
Max.PowerConsumption:1425Watts
230 VAC Unit: 230VAC,50/60Hz,6.5amp
DedicatedSupplyCircuit*:230VAC,50/60Hz,10amp
Max.PowerConsumption:1500Watts
* Power source must have over-voltage limits less than 1500 watts from mains to ground.
(Installation Category II in accordance to IEC664)
Electrical Requirements