Specifications

11
TP202 Rev. A © Midmark Corporation 2013
Installation
Operating Environment
Equipment Alert
Allow unit to reach room temperature before operating.
Failure to do so may result in damage to sterilizer.
Ambient Temperature Range: 68°F to 104°F (+20°C to +40°C)
Relative Humidity: <80%(non-condensing)
(Pollution Degree 2, in accordance to IEC664)
Normal Operating Altitude: < 9842 ft. (3000 m) above sea level
Device approved for INDOOR USE ONLY.
Device to be operated in a relatively dust-free environment.
(Pollution Degree 2, in accordance to IEC664)
Device should be connected to a power source with over-voltage limits less than 1500 watts from
mains to ground. (Installation Category II in accordance to IEC664)
The M9 and M11 will emit 5000 BTU / HR during operation.
Location Requirements
Allow Clearance
on Both Sides
2"
(5 cm)
Clearance
21"
(53 cm)
Support Surface
1"
(3 cm)
5"
(13 cm)
Requirements
5"
(13 cm)
18"
(46 cm)
M11
M9
M11
23" (58 cm)
M9
22" (56)
2"
(5 cm)
Overhang/Shelf
M9
18"
(46 cm)
M11
15"
(38 cm)