Specifications

7
DESCRIPTION
Safety Features
The M9 / M11 UltraClave™ or M9D / M11D Autoclave sterilizers have the follow-
ing Safety Features . . .
dry cycle door stop prevents door from opening fully if there is a slight resid-
ual pressure in chamber when door handle is operated.
Door stop also retains door in a partially open position during drying cycle.
door latched switch prevents operation until door is completely closed and
latched. An audible “beep” sounds and message “
DOOR OPEN, CLOSE DOOR” is
displayed if door is not fully latched.
monitors chamber temperature during a cycle to prevent an overheat con-
dition. If chamber temperature exceeds its preset limit, power to the heaters
is cut off and the cycle is aborted.
pressure relief valve opens to reduce chamber pressure should chamber
steam pressure exceed allowable limit. Released steam is directed out of
bottom of unit’s cabinet.
fault detection circuit monitors all functions of sterilizer during a cycle. If a
fault occurs during a cycle, the sterilizer monitoring circuitry stops the cycle,
sounds an audible signal, and displays an appropriate error message.
displays message “
ITEMS NOT STERILE” if a sterilization cycle is interrupted by
power loss, by the operator, or by a unit malfunction.
Description