Specifications

6
Re-Location Requirements For Sterilizer
1. Disconnect power cord from electrical outlet and allow sterilizer to cool.
2. Drain water from reservoir or do not tip sterilizer, allowing water to spill.
Electrical Requirements
Electromagnetic Interference
This Midmark Sterilizer is designed and built to minimize electromagnetic inter-
ference with other devices. However, if interference is noticed between another
device and this Sterilizer, remove interfering device from the room and / or plug
Sterilizer into an isolated circuit.
The three-pronged grounding plug on sterilizer power cord must be plugged into
a matching three-pronged, grounded, dedicated, correctly polarized receptacle.
ELECTRICAL RATINGS
M9 / M11
M9D / M11D
115 VAC Unit 115 VAC, 50/60 Hz, 12 amp
Dedicated Supply Circuit: 120 VAC, 50/60 Hz, 15 amp
Maximum Power Consumption: 1425 Watts
230 VAC Unit 230 VAC, 50/60 Hz, 6.5 amp
Dedicated Supply Circuit: 230 VAC, 50/60 Hz, 10 amp
Maximum Power Consumption: 1500 Watts
DANGER
Do not use this sterilizer in an explosive or oxygen-enriched atmo-
sphere, or where flammable anesthetics are stored. An explosion or
fire could occur, causing fatal or serious injury.
WARNING
Use 104 - 127 VAC, 50/60 HZ alternating current only for 115 VAC rat-
ed models and 207 - 253 VAC, 50/60 HZ alternating current only for
230 VAC rated models. Failure to do so could result in electrical shock to per-
sonnel and will result in damage to sterilizer.
EQUIPMENT ALERT
Unit should be allowed to reach room temperature before operating. Fail-
ure to do so could result in damage.
NOTE
Grounding reliability can only be achieved if this unit is connected to a matching three-
pronged, grounded, isolated, correctly polarized receptacle.
Installation