Specifications

5
Support Surface Depth (C) - Support surface should be at least 21 in. (53.3 cm)
(M11) or 18 in. (45.7 cm)(M9) deep. Allow at least 2 in. (5.1 cm) clearance be-
hind the sterilizer for air circulation.
Distance To Side Wall (D)
- If sterilizer is located next to side walls, there should
be no less than 2 in. (5.1 cm) clearance between the side of the sterilizer and
the wall.
Distance Above Sterilizer (E)
- If sterilizer is to be located beneath wall cabinets
or shelves, the underside of the cabinets or shelves should be at least 23 in.
(58.4 cm)(M11) or 22 in. (55.9 cm)(M9) above the support surface or *5” clear-
ance above top of sterilizer to allow access to the printer for changing the printer
paper roll.
Overhang (F)
- If the sterilizer is to be located beneath an overhang, the under-
side of the overhang should project no further than 15 in. (38.1 cm)(M9) or 18 in.
(45.7 cm)(M11) over the rear of the sterilizer.
Location On Support Surface
- The front of the sterilizer should be located no
less than 1" (2.54 cm) back, from the front of the support surface, but no more,
so water can be easily drained from the tube into a container (See sketch on
Location).
Neighboring Materials and Equipment
- If the sterilizer will be operated in con-
tinuous cycles, locate the sterilizer where steam will not damage materials or
equipment in the surrounding area.
Operating Environment Conditions
Operating Environment
Temperature Range: ................. 68°F to 104°F (20°C to 40°C)
Normal Operating Altitude: ......... Below 9842 ft. (3000 m) above sea level.
Device approved for Indoor Use Only.
Device to be operated in a relatively dust free environment. An excessive
relative humidity environment should be avoided. Less than 80% (non con-
densing) is recommended. (Pollution Degree 2, in accordance to IEC664)
Device should be connected to a power source with over-voltage limits less
than 1500 watts from mains to ground. (Installation Category II in accor-
dance to IEC 664)
The M9 and M11 will emit 5000 BTU/HR. during operation.
EQUIPMENT ALERT
The unit should be allowed to reach room temperature before operating.
Failure to do so could result in damage to the unit.
Installation