Specifications
41
When the sterilizer enters the Drying Phase the word “DRYING” is printed and
the printer will print the elapsed time in 5 minute increments starting with 0:00
and the words “Drying Start”. The printer continues to print the elapsed time in
5 minute increments until the Drying Phase is completed. The final record for
the Drying Phase will include the words “Drying Complete”. In the event the
Drying time is programmed to a time that isn’t divisible by 5 the final printed
record for the Drying Phase will reflect the actual programmed drying time in 1
minute increments, e.g. a programmed Dry time of 12 minutes will have 5, 10,
and 12 minutes printed on the printer tape.
When the sterilizer has completed the Drying Phase of the sterilization cycle the
printer will print a summary of the sterilization cycle with the duration of each
phase of the cycle and the Total Cycle Time. Following the summary the printer
will print “CYCLE COMPLETE”.
NOTE
In the event the Drying Phase is aborted by the operator before it is complete the words
“Drying Complete” and “CYCLE COMPLETE” will not be printed.