Specifications
34
SPECIFICATIONS
M9/M9D Sterilizer
Physical Dimensions:
Overall Length w/Plug .....................20.38 in. (51.8 cm)
Overall Width ...................................15.3 in. (38.9 cm)
Overall Height w/Printer ..................15.8 in. (40.1 cm)
Counter Area ...................................15.3 in. x 18 in. (38.9 cm x 45.7 cm)
Chamber 9.0 in. diam. x 15.0 in. deep
(22.9 cm x 38.0 cm)
Standard Trays ..................... Large: 7 5/16 in. x 12 in. x 7/8 in.
(18.6 cm x 30.5 cm x 2.2 cm)
Small: 5 5/8 in. x 12 in. x 7/8 in.
(14.3 cm x 30.5 cm x 2.2 cm)
Weight:
Empty Reservoir..............................73 lbs (33.1 Kgs)
With Shipping Carton ......................81 lbs (36.7 Kgs)
Water Reservoir Capacity .............1.1 Gallons (4.1 liters) to Full Mark
usable volume is 0.5 gallons (1.9 liters)
Electrical Requirements:
NOTE: A separate (dedicated) circuit is required for this sterilizer. Sterilizer
should not be connected into an electrical circuit with other appliances or equip-
ment unless circuit is rated for the additional load.
115 VAC, 50/60 Hz Model ...............15 AMP Circuit, Single Phase (1425 Watts)
230 VAC, 50/60 Hz Model ...............10 AMP Circuit, Single Phase (1500 Watts)
Fuse Ratings
115 VAC Unit
F1 .................................0.25 Amp, 250 V, Slo-blo, 1/4" x 1 1/4"
F2..................................15 Amp, 250 V, Fast Acting, 1/4" x 1 1/4"
230 VAC Unit
F1..................................0.125 Amp, 250 V, Slo-blo, 5 x 20 mm
F2..................................8 Amp, 250 V, Fast Acting, 5 x 20 mm
Chamber Pressure at
270°F (132°C) .................................27.1 psi. (186.2 kPa)
Safety Valve Setting ......................40 psi. (275.8 kPa)
Certifications .................................ASME Boiler & Pressure Vessel Code,
Section VIII, Division1.
Canadian Registration Number Available
UL61010A-1