Specifications

33
CALLING FOR SERVICE
Contact your Midmark Authorized Dealer, or log onto www.midmark.com
Model and serial number information will be required
when calling for ser-
vice. To contact Midmark directly:
1-800-Midmark (1-800-643-6275) or 937-526-3662
8:00 a.m. until 5:00 p.m. Monday through Friday
(Eastern Standard Time in the U.S.)
[excluding standard U.S. holidays].
C533 through C633
STEAM TEMP LOW
or
STEAM TEMP HARDWARE
or
PRESSURE LOW
Sterilizer detects that
temperature and / or
pressure is outside the
limits for normal
operation.
Unplug unit power cord for 1
minute and then plug back in.
If problem persists, contact an
authorized service
representative (see Calling For
Service later in this manual).
C642 through C647
PRESSURE HIGH
Pressure inside
chamber is outside the
limits for normal
operation.
Wait briefly (up to one minute)
while chamber pressure /
temperature dissipates.
Press STOP button to return to
Select Cycle mode.
Initiate a new cycle.
C660 through C677
PRESSURE HARDWARE
or
PRESSURE OVERLIM
Pressure inside
chamber is outside the
limits for normal
operation.
Unplug unit power cord for 1
minute and then plug back in.
If problem persists, contact an
authorized service
representative (see Calling For
Service later in this manual).
C980 through C987
HI-LIMIT OPEN
High Limit switch has
opened for at least 1/4
second during specific
operational mode.
Unplug unit power cord for 30
minutes ,to allow unit to cool,
and then plug back in.
If problem persists, contact an
authorized service
representative (see Calling For
Service later in this manual).
NOTE
Please mark down any displayed Code(s) and be sure to relay this information to the
Service Technician.
Message Guide - Continued
Message Possible Cause Solution
Calling For
Service
Maintenance