C060: POWER UP MODE
SYSTEM HARDWARE
Power was interupted
briefly or an internal
glitch.
Unplug unit power cord for 1
minute and then plug back in.
If problem persists, contact an
authorized service
representative (see Calling For
Service later in this manual).
C102: FILL MODE
STOP PRESSED
STOP button was
pressed during cycle.
Press STOP button to restart.
C103 through C105
HEATUP, STERILIZE, OR
VENT MODE
STOP PRESSED
STOP button was
pressed during cycle.
Wait briefly (up to one minute)
while chamber pressure /
temperature dissipates.
Press STOP button to return to
Select Cycle mode where a new
cycle may be initiated.
C106: DOOR MODE
STOP PRESSED
STOP button was
pressed during cycle.
Press STOP button to restart.
C232: FILL MODE
WATER LOW
Water level in reservoir
is too low.
Refill water reservoir with distilled
or demineralized water. Wait
briefly (up to one minute).
Press STOP button to return to
Select Cycle mode where a new
cycle may be initiated
Fill / Vent filter (in
bottom of chamber)
clogged.
Clean Fill / Vent filter (Refer to
Non-Scheduled Maintenance).
C326: DOOR MODE
DOOR CLOSED
Door latch is still
making contact after
door motor operated.
Open Door
C382 FILL
DOOR OPEN
Sterilizer detects that
door switch contacts
opened.
Close the sterilizer door.
(Cycle will continue where left
off.)
C383 HEATUP MODE
DOOR OPEN
Sterilizer detects that
door switch contacts
opened.
Wait briefly (1 minute) while
chamber pressure / temperature
dissipates.
Press STOP button to return to
Select Cycle mode.
Initiate a new cycle.
C384 STERILIZE MODE
DOOR OPEN
Sterilizer detects that
door switch contacts
opened.
Unplug unit power cord for 1
minute and then plug back in.
If problem persists, contact an
authorized service
representative (see Calling For
Service later in this manual).