Specifications
31
Message Guide
CAUTION
The Message Guide list some of the messages that may occur during op-
eration, along with possible causes and solutions.
If an error occurs more than once, do not continue to use sterilizer.
Note the message or error code, unplug unit and call an authorized service represen-
tative (see “Calling For Service”).
Also, if an error message contains the phrase “Items Not Sterile”, the items in steriliz-
er shall not be considered sterile; they must be run through a successful sterilization
cycle.
Message Guide
Message Possible Cause Solution
Informational Messages
INITIALIZING SYSTEM
Unit power cord was
just plugged in -
standard informational
message.
normal operation will occur after
a 4 second pause.
TOTAL CYCLES XXXX
M9*, vX.XX.
(*will display Model Number,
M9, M9D, M11, or MllD).
Unit power cord was
just plugged in -
standard informational
message.
normal operation will occur after
this message.
PERFORM WEEKLY
MAINTENANCE
This message is
displayed every 7, 14,
and 21 days after the
unit is plugged into a
power source to prompt
the operator to perform
weekly maintenance
described in this
manual.
Perform weekly maintenance.
The message will automatically
clear after the next cycle is
completed.
PERFORM MONTHLY
MAINTENANCE
This message is
displayed every 28 days
to prompt operator to
perform monthly
maintenance described
in this manual.
Perform monthly maintenance.
The message will automatically
clear after the next cycle is
completed.
Error Messages
C010: POWER UP MODE
SYSTEM PWR LOSS
ITEMS NOT STERILE
PUSH STOP TO RESTART
Unit had loss of power
during cycle.
Press STOP button to restart.
Maintenance