Specifications
29
Extended Use Maintenance
The M9/M9D and M11/M11D are designed and tested to provide exceptional
reliability throughout their service life. However, like all electro-mechanical
devices they are subject to wear and degradation with use.
To ensure the integrity, performance, and safety of all major components it
is the responsibility of the user to have the sterilizer performance / operation
verified by a Midmark Authorized Service Provider at least every 10 years or
10,000 cycles, whichever comes first. After 10 years or 10,000 cycles of use, an
annual inspection by a Midmark Authorized Service Provider is recommended.
Troubleshooting Guide
Use the following table to assist in correcting minor problems with the sterilizer.
TroubleShooting Guide
Problem Possible Cause Solution
Sterilizer does not
operate (no display)
Sterilizer power cord came
loose from supply outlet or
back of sterilizer.
Assure Sterilizer power cord is
plugged into outlet and
sterilizer.
No power at Sterilizer
supply outlet
Check circuit breaker for supply
outlet. If problem recurs, unplug
unit power cord and contact an
authorized service technician
(see Calling For Service).
Fuse open on main P.C.
Board.
Unplug unit power cord and
contact an authorized service
technician (see Calling For
Service).
Steam is escaping
from pressure relief
valve
pressure relief valve not
properly seated.
reseat pressure relief valve (see
Perform Pressure Relief Valve
Check under Monthly
maintenance).
Sterilization failure
evidence from
process monitor
(chemical indicator,
biological indicator,
etc.)
sterilization conditions
were not present at
location of the indicator.
reload sterilizer in accordance
with Guidelines For Loading the
M9/M9D or M11/M11D.
If problem recurs, unplug unit
power cord and contact an
authorized service technician
(see Calling For Service).
indicator is out-of-date, is
inappropriate for
sterilization cycle, or has
otherwise malfunctioned.
use an indicator, appropriate for
the load and cycle selected, that
has been stored properly.
Contact the indicator
manufacturer for additional
information on proper selection,
use, storage, and potential
misapplication or malfunction.