Specifications

28
Non-Scheduled Maintenance
1. Cleaning Fill / Vent and Air
Filter Screens
The filter screens (1) are
intended to prevent debris
from causing valve failures.
The screens should be
cleaned if fill or vent times
become too long or items
will not dry.
(a) Unplug the power cord
from outlet receptacle.
Open door and remove
trays.
(b) To access Fill / Vent fil-
ter screen,remove tray
rack.
Air filter screen is
located in back of chamber.
(c) Grasp filters (1) and gently pull upward while twisting slightly
(a pair of pliers may be used if filter is stuck).
Clean filters with mild soap and distilled water. A small stiff bristled
brush or ultrasonic cleaner may be helpful. Rinse filters with distilled
water.
(d) Re-install filters (1) by inserting into the hole, and pressing downward
while twisting slightly.
(e) Install tray rack.
(f) Plug unit power cord into outlet receptacle.
CAUTION
Always unplug unit power cord from outlet and allow unit to cool be-
fore performing maintenance. Failure to comply may result in injury.
EQUIPMENT ALERT
Do not operate sterilizer without filters in place.
MA601400i
1
Maintenance