Specifications
25
2. Installation
• Place back edge of tray in chamber.
Pressing downward (A) on top of tray
rack, slowly insert assembly into
chamber (B).
Daily
1. Clean External Surfaces .
2. According to your facility’s procedure
:
• Use only quaternary disinfectants to disinfect unit . Staining, pitting, dis-
coloration, or softening could occur if phenolic, iodophor, or glutaralde-
hyde-based disinfectant is used on plastic surfaces of the unit. Also, use
of alcohol or aerosol spray cleaner / disinfectant containing substantial
amounts of alcohol in the formula can damage the faceplate.
• Wring excess solution from cloth.
• Using soft cloth, wipe all external surfaces.
• Do not rinse or dry external surfaces. Allow germicidal solution to air dry.
3. Clean Sterilizer Door Gasket
• Clean door gasket sealing lip and mating surface with a damp cloth.
• Examine gasket for possible damage.
EQUIPMENT
ALERT
Install tray rack and plate assembly
so that angled end of plate is toward back
of
chamber. Do not
allow plate to contact water
lever sensor.
EQUIPMENT ALERT
If the sterilizer is used frequently to process dental handpieces that have
been lubricated or dipped in dental milks, drain the water from the reservoir
daily. Refill the reservoir with distilled water.
CAUTION
Make sure that unit is cool when cleaning door gasket and any mat-
ing surfaces to prevent being burned.
MA601600i
B
A
Maintenance