Specifications
21
then displays: Press
Displays: then
Press changes VENT to FAST
. Press changes VENT to SLOW.
When in FAST
vent, valve fully opens and vents chamber.
When in SLOW
vent, valve opens periodically for a fraction of a second once
per minute to slowly vent chamber.
then displays: Press
Displays: then
Press raises time by 1 minute. Press lowers time by 1 minute.
Dry time can be adjusted from zero
to a maximum of 60 minutes, in 1 minute
increments.
then displays: Press
The display shows the new programmed conditions.
Displays: then
Load articles to be sterilized into chamber.
Press START button to begin sterilization cycle for Program 1.
NOTE
The programmed settings entered are retained under Program 1 button. Even if power
is interupted or the unit unplugged the settings will be retained.
CAUTION
All materials run in these cycles must be validated by the user.
Failure to do so could result in incomplete sterilization.
NOTE
After any cycle has been selected but not started, depressing the STOP button
or allowing 5 minutes or more time to pass, the display will go back to
.
<P> FOR NEXT
<STOP> TO CANCEL
VENT TYPE:
FAST
<+> OR <
-
> TO ADJUST
<P> FOR NEXT
<STOP> TO CANCEL
DRY TIME
XX MINUTES
<+> OR <
-
> TO ADJUST
<P> FOR NEXT
<STOP> TO CANCEL
PROGRAM 1
XXX°F, XX MINUTES
FAST VENT
XX MINUTE DRY
SELECT CYCLE
Operation