Specifications
20
Programming Mode
The following steps are for programming buttons 1 and/or 2 for applications that
are not covered by standard cycle programs:
Displays: Press
Displays: then Press
Displays: then then
Press raises temperature 1°. Press lowers temperature 1°.
Sterilization temperature can be adjusted from a minimum of 230°F (110°C)
to a
maximum of 275°F (135°C).
Permitted temperature range for proper sterilization
is 250°- 275°F (121°- 135°C).
then displays: Press
Displays: then
Press raises time 1 minute. Press lowers time 1 minute.
Sterilization time can be adjusted from a minimum of 3 minutes
to a maximum of
90 minutes. It is important to properly coordinate the cycle time with the steriliza-
tion temperature.
CAUTION
Temperatures set below 250° F (121°C) should not be used for steril-
ization, unless otherwise required by the device manufacturer. Tem-
peratures below 250° (121°C) are provided for disinfection only.
NOTE
If the STOP button is pressed anytime during the Programming Mode the settings en-
tered will be aborted and revert back to the original settings before programming be-
gan.
MA596700i
SELECT CYCLE
PROGRAM 1
XXX°F, XX MINUTES
FAST VENT
XXX MINUTE DRY
STERILIZATION TEMP:
XXX.X°F
<+> OR <
-
> TO ADJUST
<P> FOR NEXT
<STOP> TO CANCEL
<P> FOR NEXT
<STOP> TO CANCEL
STERILIZATION TIME
XX MINUTES
<+> OR <
-
> TO ADJUST
Operation