Specifications

19
Automatic Door Open (Pertains only to M9/M11 UltraClave™)
Door To Open
Display shows:
An Audible signal is emitted to indicate to Operator door is about to open.
When pressure in chamber reaches zero, door actuates to partially open
(Drying mode) position.
Manual Door Open
(Pertains only to M9D/M11D AutoClave)
Open Door
Display shows (top display line flashes):
An Audible signal is emitted when pressure inside chambers reaches zero
to indicate to Operator to open door. The door should be opened to the
first stop (drying mode) position.
The audible signal will continue to repeat every minute until either the
door is opened to the DRY (partially opened) position, or by pressing the
STOP button, aborting the DRY cycle.
Drying
Display shows:
Time of Dry Cycle is counted down.
Dry Cycle Complete
when Drying time reaches 0:00.
Display shows:
An audible signal is emitted for 10 seconds, the display shows:
4. Remove processed load from chamber (See Unloading Hot Trays and Cas-
settes later in this manual).
5. The sterilizer may now be reloaded for another cycle.
NOTE
If desired, Drying Cycle can be aborted by pressing STOP button.
CAUTION
The processed load and inner surfaces will be hot. Avoid contact
with hot surfaces. Failure to do so could result in serious burns
DOOR TO OPEN
XXX°F XX.X PSI
OPEN DOOR TO DRY
STOP TO ABORT
DRYING
MM:SS
DRY CYCLE
COMPLETE
SELECT CYCLE
Operation