Specifications
18
Press START.
Sterilizer sounds a “beep
” for two seconds to indicate cycle has started.
• Filling Chamber
:
Display shows:
Chamber automatically begins to fill to the correct level with water.
When chamber is full
, display shows
• Heat Up
portion of the cycle begins.
Display shows:
(can also be in metric values depending on set-up).
Display changes as temperature and pressure in chamber changes.
• Sterilizing
portion of cycle begins when correct temperature and pres-
sure is reached.
Display shows:
Time remaining in cycle is counted down while current temperature and
pressure in chamber is continuously updated.
• Ready to Vent
is displayed when there is 10 seconds remaining in Steril-
ization Cycle.
Display shows:
• Vent opens
when time runs out in Sterilizing mode.
Display shows:
The display changes as temperature and pressure in chamber changes.
Vent valve opens, venting steam and water from chamber back into reser-
voir.
WARNING
STOP button may be pressed at any time to stop or interrupt a cycle.
Goods must not
be considered sterile if this occurs before Dry Cycle.
Sterilizer will return to SELECT CYCLE mode.
CAUTION
Keep clear when M9/M11 door is ready to open. Do not attempt to
open M9D/M11D door until steam dissipates.
Failure to do so could result in severe burns from steam being released.
FILLING
CHAMBER
CHAMBER
IS FULL
HEATING - UNWRAPPED
XXX°F XX.X PSI
STERILIZING
MM:SS XXX°F XX.X PSI
READY TO VENT
XXX°F XX.X PSI
FAST VENT
XXX°F XX.X PSI
Operation