Specifications
17
Standard Cycle Operation
Refer to following steps for a detailed description of operating procedures:
1. Display shows:
2. Close and latch door; sterilizer will not
operate unless door is closed and
latched properly.
3. Select either Unwrapped, Pouches, Packs, or Handpieces sterilization
cycle in accordance with the Standard Cycle Parameters table.
• Press UNWRAPPED. Display shows the program’s parameters:
and then
EQUIPMENT ALERT
Using an incorrect sterilization program could result in non-sterile goods
and may damage instruments. Consult instrument manufacturer for spe-
cific sterilization instructions.
NOTE
Pressing the START button when SELECT CYCLE is displayed, at beginning or end of
a cycle, activates the heater for 10 minutes
. The display flashes
ADDITIONAL HEAT. This allows the Operator to preheat the chamber before starting
an operation or to add additional time to the Dry mode at the end of an operation.
Pressing STOP will end the ADDITIONAL HEAT time.
NOTE
On units using the metric display,°F will display as°C and PSI will display as KPA.
Pressing enables operator to change DRY time from 0 to 60 minutes in 1 minute
increments on a pre-programmed cycle.
Pressing decreases time. Pressing increases time.
MA596700i
SELECT CYCLE
UNWRAPPED
270°F, 3 MINUTES
FAST VENT
30 MINUTE DRY
Operation