Specifications
15
† Packs to have a minimum of 1/4 in. (6.4 mm) space between each other and
away from all sterilizer surfaces.
Operating the Sterilizer
Packs (†)
4425 cu. cm up to 5.0
cm thick (270 cu. in.
up to 2 in. thick)
3195 cu. cm up to 5.0
cm thick (195 cu. in.
up to 2 in. thick)
15240 cu. cm up to 5.0
cm thick (930 cu. in. up
to 2 in. thick)
WARNING
Do not use this sterilizer for sterilizing volatile substances or for any
purpose other than its intended design. Burns and toxic or explo-
sive conditions could result.
Do not
force door handle at any time. Chamber pressure may cause door to
open with extreme force. If door handle does not move freely, allow unit to cool
and depressurize for 40 minutes before opening door. Failure to adhere could
result in serious personal injury.
CAUTION
Programmable cycles 1 & 2 are provided for those applications requiring
sterilization parameters different than the preset cycles. All material pro-
cessed in these cycles must be validated by the user
to ensure sterility of the pro-
cessed load.
Maximum Capacities
Load Type
M11/M11D
Large / Deep
Tray
M11/M11D
Small Tray
M11/M11D
Sterilizer Total
Operation