Specifications

14
Place all containers so that the
opening allows steam to enter,
and air and condensate to
drain from the container.
Sterilize jointed items in an
open position.
Use only M9/M9D or M11/M11D trays in their appropriate sterilizer. Using
other trays could restrict air and steam flow to items.
Place unwrapped items on a towel or absorbent paper.
Pouches may overlap, but handpieces and instruments must be single
height loaded (not piled or stacked), to permit proper steam flow and pene-
tration to the items.
Refer to below chart for maximum M9/M9D sterilization loads. If a load sur-
passes these limits, we recommend dividing the load and running multiple
cycles. Listed below are the maximum recommended loads for each
tray
and total
loads:
Packs to have a minimum of 1/4 in. (6.4 mm) space between each other and
away from all sterilizer surfaces.
Refer to below chart for maximum M11/M11D sterilization loads. If a load sur-
passes these limits, we recommend dividing the load and running multiple cy-
cles. Listed below are the maximum recommended loads for each
tray and total
loads:
Maximum Capacities
Load Type
M9/M9D Large /
Deep Tray
M9/M9D Small
Tray
M9/M9D Sterilizer
Total
Solid Items 42 instruments - 1089
grams (2.4 lbs.) or
28 instruments - 726
grams (1.6 lbs.) or
140 instruments - 3629
grams (8.0 lbs) or
Handpieces 9 in rack or 9 in rack or 9 handpieces in rack
and 75 instruments or
Packs
(†)
1082 cu. cm up to 2.5
cm thick (66 cu. in. up
to 1 in. thick)
787 cu. cm up to 2.5
cm thick (48 cu. in. up
to 1 in. thick)
3736 cu.cm up to 2.5
cm thick (228 cu. in. up
to 1 in. thick)
Maximum Capacities
Load Type
M11/M11D
Large / Deep
Tray
M11/M11D
Small Tray
M11/M11D
Sterilizer Total
Solid Items 45 instruments - 1225
grams (2.7 lbs.) or
30 instruments - 816
grams (1.8 lbs.) or
150 instruments - 4082
grams (9.0 lbs) or
Handpieces 9 in rack or 9 in rack or 9 handpieces in rack
and 75 instruments or
Operation