User Manual

7
M3 Sterilizer - continued
Note
Clearance.
Maintain a minimum of 6 inch clearance above the condensing tank for
proper steam ventilation. If enclosed in a cabinet the support surface
and surrounding surfaces, should be protected with a water resistant
material (e.g. plastic, laminate, stainless steel, etc.). If enclosed in a
cabinet, it is recommended that the door be vented to avoid heat,
moisture build up and potential damage to the inside of cabinet.
Connect the condensing tank - continued.
F) Place tank on a level surface, preferably below the sterilizer
but in no case should it be higher than the sterilizer support surface.
G) Connect sensor plug.
H) Plug power cord into receptacle on back of sterilizer.
I) Plug power cord into a properly polarized and grounded receptacle
rated for a minimum of 15 amps. A dedicated circuit only used for
the sterilizer is recommended.
0
0
Power Cord
Sensor Plug