User Manual
6
M3 Sterilizer - continued
0
0
Equipment Alert
The tank hose must be properly connected
(and not kinked). Improper connection will
cause water / steam leaks or a sterilizer
malfunction if water flow to the tank is restricted.
Note
External Condensing Tank is not needed if M3 is connected
to the VistaCool direct-to-drain thermal reduction system.
Connect the condensing tank.
A) If connected, disconnect sterilizer power cord.
B) Push tank hose into compression fitting on lid.
Tighten nut. (If step B is difficult - see NOTE)
C) Connect tank hose to back of sterilizer.
D) Fill condensing tank with tap water to minimum
water level indicator line.
E) Secure the lid on the condensing tank by aligning
the tabs and rotate lid clockwise to secure.
Note
If tubing is difficult to install.
A) Remove compression nut & two ferrules.
B) Install nut and ferrules onto tank hose as shown.
(Large & small ferrules must be positioned as shown)
C) Insert hose into tank fitting, then tighten nut.
Power Cord
Compression
Fitting
Tank Hose
Min. Water Level
Indicator Line
Ferrule
(small)
Ferrule
(large)
Nut