User Manual

3
Voltage Rating
Connect power cord.
A) Plug power cord into receptacle on back of sterilizer.
B) Plug power cord into a properly polarized and grounded receptacle
rated for a minimum of 15 amps. A dedicated circuit only used for
the sterilizer is recommended.
C) M9 / M9D and M11 are not equipped with an on/off switch, the display
operates off very low power. (example: microwave oven display)
Note: When power is connected, the messages shown will appear on the display.
Fill reservoir.
A) Open door to unit.
B) Pour distilled water into fill opening until
water level reaches the top of the fill level
label on the water level indicator tube.
Equipment Alert
Use distilled water or water that meets the
water purity specifications referenced in
your Users Guide. Failure to comply may
result in sterilizer malfunction and/or premature
failure due to excessive corrosion.
*
*
M9 / M11 Steam Sterilizer - continued
Note
For M9-04x / M11-04x models, The User Settings mode will enable
the user to set the desired unit of measure and adjust the time clock.
Refer to “User Settings” in your User Guide.
M9-04x / M11-04x
Display Shown
* These screens will display the total number of cycles run on the unit, the
model number (M9/M9D or M11), the software version number, serial number
and M9/M11-04X models also display the date and time.