Once the required number of Post-BD tests is performed, click the Save Review button
to display the Pre/Post comparison.
VC or SVC
This test can determine if a patient’s lungs are trapping air during a forced maneuver
(FVC test). Air trapping can be an indication of airway obstruction and is also sometimes
seen in older patients. This is a very slow and deliberate test. The patient takes 2 or 3
normal tidal breaths and then a slow, deep breath, and then a slow, full exhale.
Click the VC button. The VC button appears bold. If this is a Pre-BD test, make sure the
Pre message is displayed in the blue box in the middle of the test screen. Do not click the
Start New Test button until the patient is ready to perform the test.
To perform a VC test, the patient must perform the test according to the following
instructions:
Inhale completely before placing the flow sensor mouthpiece into the mouth.
Insert the mouthpiece just past the front teeth and seal the mouth to the flow
sensor mouthpiece with the lips.
Take 3 or 4 tidal breaths with the mouthpiece in the mouth.
Inhale as deeply as possible, until the technician observes a plateau at the
bottom of the graph.
Blow out slowly and evenly until the technician observes a plateau at the top
of the graph or the device signals that the end of test criteria has been met.
The technician should stop the test with the mouse or the ENTER key.
Remove the mouthpiece from their mouth when instructed by the technician.
The patient must exhale at a faster rate during the “Blow out” phase than during the tidal
breaths. After the patient has completed the number of tests you require, click the Save
Review button and click Yes to save the test efforts.
MVV
The Maximum Voluntary Ventilation, or MVV, is the measurement of a patient’s
breathing when the patient inhales and exhales maximally and rapidly for 12 to 15
seconds. The software takes this result and extrapolates the results for a period of one
minute. MVV is expressed in liters per minute (L/min). This is a very demanding test
and patients must be allowed to rest between efforts. The MVV test is required to qualify
some patients for Social Security Disability benefits.
Click the MVV button. The MVV button now appears bold. For a Pre-BD test, make
sure the Pre message is displayed in the blue box in the middle of the test screen. Do not
click the Start New Test button until the patient is ready to perform the test.