Spirometry Settings Calibration Tab
Select the Calibration Reminder box to remind the user when the Spirometer has not
been calibrated for the interval set in the Calibration Expiration Interval box.
Starting a New Calibration
Click New Cal on the Calibration screen to begin a new calibration. Attach the 3-liter
syringe to the Calibration Adapter, and then attach the adapter to the mouthpiece. Check
that the mouthpiece is properly attached to the handle. The first time you calibrate your
Spirometry handle, you must type your name in the Performed by box, the Syringe SN
(serial number) and Spirometer SN (handle serial number) in the appropriate boxes.
Click Start Cal. The sensor will zero itself and then the following screen appears.