Service manual

IQmark Digital Spirometer Operation Manual
27
E.
Calibration
The ATS recommends daily calibration of any Spirometer. The IQmark Digital
Spirometer automates this process for quick and accurate preparation of the instrument.
Calibrating the Spirometer requires a 3-Liter syringe. Midmark Diagnostics Group
strongly recommends a Midmark Diagnostics Group 3-Liter syringe with the Midmark
Diagnostics Group Calibration Adapter.
NOTE: The Spirometer handle must be calibrated with a Disposable Pneumotach
Mouthpiece (DPM). The DPM must be calibrated with a Calibration
Adapter that fits OVER the outside diameter of the mouthpiece. Never
calibrate the device with a syringe that fits within the DPM.
You can access the Calibration program from the Windows Task Bar through the
following sequence:
Start Programs Diagnostic Workstation Spirometry Calibration
or Double-click the Spirometry Calibration icon:
The following screen is displayed. Click the handle icon in the upper left corner to start
calibration. Clicking OK will close this screen and return to the desktop. It is
recommended that you place the DPM into the handle before you enter the Calibration or
Test screen.
Spirometry Calibration Utility Screen
Next
Back
Go To Table Of Contents