Service manual
IQmark Digital Spirometer Operation Manual
20
Interpretation Tab
From this tab, you can select which Reference Equations set to use as the default for
Adults and Pediatrics. Not every Spirometry measurement (FVC, FEV
1
, etc) are
calculated by every Reference Equation. The Secondary reference equations can be used
to “fill in” missing measurements. For example, if your primary reference equation does
not have an equation for FEF
25
, you can select a secondary reference equation that does
to “fill in” your report.
Spirometry Settings – Interpretation Tab
You can also turn on or off the Auto. Interpretation feature. If the Auto. Interpretation
box is selected, you can select the ATS Logic or the NHANES III logic. If you do not
wish to have the software automatically interpret the Spirometry tests, clear the Auto.
Interpretation box. The Primary Care Practitioner (PCP) Mode overrides Reference
Equation and Interpretation selection. It is discussed on the next page.
Next
Back
Go To Table Of Contents