(1) For the greatest accuracy, calibrate the Spirometer frequently. Daily calibration is
recommended by the ATS.
(2) DO NOT RE-USE Disposable Pneumotach Mouthpieces.
(3) Replace the batteries when the device LED turns Red or the computer screen
displays the “low battery” indication.
(4) Inspect the Spirometer periodically for accuracy. Contact Midmark Diagnostics
Group for your annual calibration certifications.
Troubleshooting
Problem Cause Solution
Volume numbers appear
too high or too low
1. Ambient air pressure or
temperature has changed
since the last calibration.
2. The current lot of
pneumotach is slightly
different from the last lot.
3. The “Use Dry Air
Conditions” option was
checked.
1. Calibrate the Spirometer.
2. Calibrate the Spirometer.
3. De-select the “Use Dry Air
Conditions” option when
performing patient tests.
This option disables the
BTPS corrections.
FVC Test Starts for no
apparent reason
1. The minimum flow
required to start a test has
been exceeded.
2. Excessive electromagnetic
interference is affecting
the instrument.
1. Instruct the patient not to
move the handle abruptly,
which can cause air to flow
through the pneumotach
and start a test.
Turn off fans or close
heating and air
conditioning vents near the
test area. Close doors or
windows.
Delay pressing the “Start
New Test” button until just
prior to beginning the
maneuver.
2. Select a different location
or find and correct the
source of the interference.
Contact Midmark
Diagnostics Group
Customer Service for
additional assistance.