Specifications
IQmark Digital ECG Operation Manual
38
C. Maintenance and Storage of the ECG Module
Preventative Inspection
A preventative inspection should occur prior to each use of the ECG module to verify
that there is no visible damage to the unit that may cause it to malfunction.
Visual inspection should include the module and all cables for signs of damage and
deterioration, including but not limited to cracks, cuts, discoloration or oxidation. If a
cable or other accessory exhibits any of these symptoms, replace it prior to using the
ECG module.
Cleaning
Clean the outside of the ECG module with a mild solution of detergent and water using a
soft cloth. Avoid using excessive amounts of solution, which may infiltrate the
connectors, battery compartments or ECG module. If necessary, use a mild sterilizing
detergent with low alcohol content, such as those generally used in hospitals. Verify
that all equipment, including accessories, are completely dry before using.
CAUTION: Do not use aromatic hydrocarbons, rubbing alcohol, or chlorinated
solvents for cleaning the ECG module.
Storage
To prevent damage to the IQmark Digital ECG module due to battery leakage or
oxidation, remove all batteries if the IQmark Digital ECG module is not to be used for a
long period of time. Avoid extreme humidity and heat during storage.
CAUTION: To prevent damage to the cables, do not hang the IQmark ECG module by
the interface cable or the patient cable.
D. Maintenance and Storage of 10-Lead Resting ECG Patient Cable
Instructions for Use
Check the cable integrity before each use. In case of damage of any kind, do not use
and do not attempt to repair cable. Contact Midmark Support Services for a
replacement cable. If cable is found to be contaminated, clean and disinfect it
according to instructions below before reusing.
Plug the 10-Lead Resting ECG Patient Cable to the ECG module as described in the
IQmark Digital ECG Operation Manual. Make sure the connection is tight. Check the
connection before each use.
Plug the metal post of each lead of the patient cable into an ECG clip. Make sure
each ECG clip is pushed all the way in.
Go To Table Of Contents